U3022AE10
Controlling the Test Set with N5222A/B, N5232A/B or N5242A/B
Sweep Setup for Multiport and Standalone Modes
When the test set is connected to the network analyzer, it is recommended that the
analyzer’s Sweep Setup be configured to Stepped Sweep before calibrating. This is
slower than the Analog Sweep, but is more accurate due to the extra electrical length
of the test set and test port cables. Stepped Sweep is available on all models.
1.
On the analyzer select
STIMUL
U
S
>
Sweep
>
Sweep Setup
.
2.
Select
Stepped Sweep
(
sweep moves in discrete steps
)
.
3.
Set the
Dwell Time
to
5
µ
s
>
OK
.
Figure 3
3
Sweep Setup
1. S
elect
[Sweep] > Sweep Setup > Timing
.
2.
Select
Timing
tab
.
3.
Se
lect
Sweep Mode: Stepped.
4. Deselect
Time: Auto Sweep Time.
5. Set
Dwell Time > Apply > OK.
Figure 3
4
Sweep Setup
, N52xxB
Sweep Setup for PNA-X N52xxB Models
Keysight U3022AE10 User's and Service Guide
41