U3022AE10
Controlling the Test Set with N5222A/B, N5232A/B or N5242A/B
Trace Measure S-Parameter
S-Parameter selection can be accomplished using
Response > Measure
. Use the
drop-down menu to select 1 of 144 S-Parameters for the 12-Port system. The first
number in the Sxx selection is the Receiver Port and the second number will be the
Source Port. Any port can be selected to be the Receiver, Source or both, as in S11.
The front panel R LED indicates the port is the Receiver and the S LED indicates the
port is the Source.
Figure
29
S-Parameter Measurement
Figure 3
0
New Trace Measure S-Parameter
S-Parameter Tab:
Multiple S-Parameters can be made from the
New Trace
menu. In
the drop-down menu select
Trace/Chan
>
Trace
>
New Trace
-OR-
(Instrument > Trace > New Trace).
The dialog box allows
the selection of any of the
144 S-Parameters.
New S-Parameter Measurement
Keysight U3022AE10 User's and Service Guide
39