This first command creates a new S10_10 measurement on channel 1
of the PNA-X, named “ch1_S10_10,” and configures the internal Test Set RF
switches.
CALC1:PAR:EXT 'ch1_S10_10', 'S10_10'
This second command feeds the newly created measurement named
“ch1_S10_10” to trace 2 on the PNA-X so that it will be displayed on the
PNA-X screen.
DISP:WIND1:TRAC2:FEED 'ch1_S10_10'
NOTE: Here are syntax format examples for single digit S-parameters:
‘ch1_S99’
‘ch1_S9_10’ ‘ch1_S10_9’
‘ch1_S22’
Method 2 - Using the Test Set Address and Data Values
This method is available while the system is in Standalone mode only.
The Address and Data values for the Test Set can be found in the
“Address and Data Values” section on page
The Test Set internal switch settings are programmed into the firmware of
the Test Set and can be controlled with the specific Address and Data
values. Each Address and Data value pair sets the Source or Receiver
switch paths. For an S-parameter measurement, two pairs of Address and
Data values will be needed.
Address and Data values are separated by a comma. Commands should
be separated by a new line, or carriage return.
For example:
CONT:EXT:TEST:DATA <address>,<data>
CONT:EXT:TEST:DATA 0,0
Example: CONT:EXT:TEST:DATA 0,0
Figure 4
4
Method 2 - Using Test Set Address and Data Values
U3022AE10
Controlling the Test Set with N5222A/B, N5232A/B or N5242A/B
Keysight U3022AE10 User's and Service Guide
51