U3022AE10
Address and Data Values
Address and Data Values
Setting the Test Port Paths with Address and Data
Example 1:
Example 2:
Table 1
1
Refer to Table 1
1
below
and
for information to set the internal
switch paths of the test set.
The address is the first value in the Test Set I/O control or GPIB data command. The
second value controls the source and receiver paths of the ports. To select a test set
port configuration both switches must be set to complete the source and receiver
paths. To do this you must add the data values together to determine the data
command value for each port, which is the second number in the Test Set I/O or GPIB
command. Refer to
“Interface Control Mode” on page 4
.
If the ports have the same address, only one command is needed.
Port 3 is the Source and Port 11 is the Receiver.
Source Port 3 = address 0, data 1 and Receiver Port 11 = address 0, data 80. The data
values are added together, the entry will be 0.81.
If the ports have different addresses, two separate address data commands must be
used.
Port 7 is the Source and Port 8 is the Receiver
Source Port 7 = address 0, data 3 and Receiver Port 8 = address 16, data 48.
Two separate commands must to be sent, you may use the same dialog box. The
entry will be 0.3 in one command line, and 16.48 in the second line.
Test Port Address and Data Values
Address
Source Path
Receiver Path
Data
0
1
2
3
4
5
0
16
32
48
64
80
0
Ports
1
3
5
7
9
11
1
3
5
7
9
11
16
Ports
2
4
6
8
10
12
2
4
6
8
10
12
52