Cal Kit Operational Check
Cal
Set Verification
1.
Select
Trace/Chan > Trace > Delete Trace
-OR-
(Instrument > Trace > Delete Trace)
. There should be no traces on the
screen.
2.
To launch the Cal Set Viewer toolbar. Select
Response > Cal > Manage Cals > Cal
Set Viewer
-OR-
(Response > Cal > Cal Sets & Cal Kits > Cal Set Viewer [On/Off]).
Figure
5
1
Calibration Cal Set Viewer
3.
From the Cal Sets drop-down menu, select
999.1
and select
Enable
. Select the
Reflection Tracking(x,x),
where x,x is the port being tested. Ensure that the
Enable
and
Error Terms
check boxes are selected.
4.
Compare the Reflection Tracking (1,1) trace to the appropriate limits in
. This can be done using Markers.
5.
Repeat step 3 and step 4 for Cal Sets "999.1" thru "999.12" (12-Port).
64
Keysight U3022AE10 User's and Service Guide