U3022AE10
N5222A
/B
and N5242A
/B
RF Interface Cable Connections
N5222A
/B
and N5242A
/B
RF Interface Cable Connections
illustrates the cable configuration of the test set to the analyzer. The cables
have been supplied with Cable Kit (U3021-60016).
1.
Remove the SOURCE OUT to CPLR THRU and RCVR IN to CPLR ARM jumpers on
the analyzer. The RCVR R1-R4 to SOURCE OUT reference loop jumpers remain on
the front panel.
2.
Add the front panel port labels to the analyzer. Refer to
.
• Port 13 over Port 3 (U3022-80002)
• Port 14 over Port 4 (U3022-80003)
3.
Connect the RF interconnect cables from the analyzer to the test set in the order
listed. As you are connecting each cable, torque to 8 in-lb. The longer, straight
end of each cable is connected to the test set. Refer to Table
9
below and
Each end of the interconnect RF cables have a different length from the bend. When
connecting the RF Interconnect cables be sure that the longer end from the bend is
connected to the PNA.
Over torque will cause damage to the test set and may cause connectors to spin or
become loose.
Table
9
RF Interface Cable Connections (U3021-60016)
Numeric
Order
RF Cables
From: Analyzer
To: Test Set
1
U3042-20042
Port 1, CPLR ARM
CPLR ARM
2
U3042-20043
Port 1, SOURCE OUT
SOURCE OUT
3
U3042-20044
Port 1, RCVR A IN
A IN
4
U3042-20045
Port 1, CPLR THRU
CPLR THRU
5
U3042-20046
Port 2, CPLR THRU
CPLR THRU
6
U3042-20047
Port 2, RCVR B IN
B IN
7
U3042-20048
Port 2, SOURCE OUT
SOURCE OUT
8
U3042-20049
Port 2, CPLR ARM
CPLR ARM
Keysight U3022AE10 User's and Service Guide
29