U3022AE10
Controlling the Test Set with N5222A/B, N5232A/B or N5242A/B
RF Path Configuration with Option 029
If the analyzer has Option 029 or H29 (noise figure capability) verify that the path
configuration is set to Default. If it is not in Default configuration, ports 5 and 9 will not
operate correctly. Select
Trace/Chan > Channel
>
Hard
ware Setup > Path Con
fi
g
...
-OR-
(Instrument > Setup > Internal Hardware > RF Path Config...)
and in the drop-down
menu select
Default
>
OK
. Save this configuration as a User Preset by selecting
Save
>
User Preset
>
Save current state as User Preset
.
Do not use the factory Preset
(User Preset Off)
;
the analyzer will return to Option 029 path configuration.
Figure 3
5
RF Path Configuration
42
Keysight U3022AE10 User's and Service Guide