Key Reference
MODE
142
Reflection Measurement
The reflection measurement mode provides the reflection
measurement, one port cable loss and distance to fault
measurement function.
Auto Scale
Press this softkey to adjust the amplitude scale automatically to
include the peak and valley level in graticule.
Key access:
Amplitude > Auto Scale
Calibrate
This softkey activates the calibration process for the measurement.
Key access:
Frequency > Calibrate
Cal Type
This softkey toggles the calibration type between full span
calibration and selected span calibration.
Full span calibration runs an open-short-load calibration across
the full frequency span.
A FullSpan calibration ignores the frequency changes during
measurements.
Selected span calibration performs an open-short-load calibration
across a user-defined frequency span. A SelectedSpan Cal requires
a re-calibration of the analyzer, after changing the measurement
frequency range. Normally, the selected span should be set as
close to the actual swept frequency range.
Key access:
Freq > Cal Type
Run/Hold
This softkey toggles the frequency sweep between run and pause
status.
Key access:
Sweep > Run/Hold