Key Reference
Sweep
160
In zero span
Auto-coupled does not function when in a zero span. The sweep
time can only be set manually. The sweep time is selectable from 6
s to 200 s in zero span.
In FFT Mode
In FFT mode, the sweep time default is auto-coupled. The
submenu, sweep time is invalid in FFT mode.
Key access:
Sweep
Sweep
Single Cont
Pressing
Sweep Single
puts the analyzer in single-sweep mode.
Press
Sweep Single
to initiate a sweep when in single-sweep mode.
Press
Sweep Cont
Performs one sweep after another as soon as it is
triggered. Pressing
Preset
(When
Preset Type
is set to
Factory
) sets
the sweep mode to continuous sweep.
Key access:
Sweep
Single Sweep
When the analyzer is in continuous sweep mode and not in a
measurement mode
(MEAS > Measure Off
), pressing
Sweep >
Single Sweep
will change the sweep setting to single sweep. It will
then begin a sweep after the trigger conditions are met. If the
analyzer is already in single sweep, pressing
Sweep > Single Sweep
will execute a new sweep as soon as the trigger condition is met.
If average is on
(TRACE > More > Average Number On
), pressing
Sweep > Single Sweep
will halt averaging of the current value, and
then continue averaging from the current count number when a
single sweep function triggers.
Key access:
Sweep
Summary of Contents for N9321C
Page 36: ...Getting Started Helpful Tips 28 ...
Page 110: ...Functions and Measurements Files Operation 102 ...
Page 180: ...Key Reference Trace Det 172 ...
Page 212: ...Menu Maps Name Editor Menu 204 Name Editor Menu ABCDEFG HIJKLMN Return OPQRSTU VWXYZ Name ...
Page 214: ...Menu Maps SPAN Menu 206 SPAN Menu Span Span Full Span Zero Span Last Span ...
Page 216: ...Menu Maps Trace Det Menu 208 ...
Page 219: ......