Functions and Measurements
Analysis the Modulated Signals
92
3 Press
Frequency > Carrier Freq > 100 > MHz
to set the carrier
frequency.
4 Press
Meas > Symbol Rate > 10 > ksps
to set the symbol rate.
5 Press
Filter Setup > Ref Filter > Nyquist
to set the filter type.
6 Set the attenuation state to Auto (default):
Press
Amplitude > Attenuation Auto
7 Setting the FSK deviation limit and turn on the limit:
Press
Shift > Limit > FreqDev UP > 85 > kHz > FreqDev Low > 80 >
kHz > Limits on
The figure below shows the FSK symbol and the measurement
result.
Figure 3-32 FSK Symbol View