Key Reference
Meas
126
• Setting dBc
Press
{OBW} > {Method} > {dBc}
to specify the power level used to
determine the emission bandwidth as the number of dB down from
the highest signal peak, within the occupied bandwidth span. This
function is an independent calculation from the OBW calculation.
The x dB bandwidth result is also called the emissions bandwidth
(EBW). This sets the green frame (f1 and f2 are the two cross points
of the signal and the pair of vertical lines of the displayed green
frame) to your specified dB value below the maximum power, and
compute the total power in the frame. The emission bandwidth is
then calculated ranging from 0.1 dB to 100 dB (the default value is
set to 26 dB).