Functions and Measurements
Making Distortion Measurements
54
The signal produces harmonic products (spaced 200 MHz from the
original 200 MHz signal) in the analyzer input mixer as shown in
.
4
Change the center frequency to the value of the first harmonic:
Press
Peak Search > Next Peak
Press
Marker > Marker To > To Center
5
Change the span to 50 MHz and re-center the signal:
Press
Span > 50 > MHz
Press
Peak Search
Press
Marker > Marker To > To Center
6
Set the attenuation to 0 dB:
Press
Amplitude > Attenuation > 0 > dB
7
To determine whether the analyzer generates harmonic distortion
products, first display the trace data in trace 2 as follows:
Press
Trace/Det > Select Trace > Trace 2
Press
Clear Write
8
Allow trace 2 to update (minimum two sweeps), then store the data
from trace 2 and place a delta marker on the harmonic of trace 2:
Press
Trace/Det > View
Press
Peak Search
Press
Marker > Mode > Delta
The analyzer display shows the stored data in trace 2 and the
measured data in trace 1. The
Marker
indicator reads the
difference in amplitude between the reference and active markers.
9
Increase the RF attenuation to 10 dB:
Press
Amplitude > Attenuation > 10 > dB
Summary of Contents for N9321C
Page 36: ...Getting Started Helpful Tips 28 ...
Page 110: ...Functions and Measurements Files Operation 102 ...
Page 180: ...Key Reference Trace Det 172 ...
Page 212: ...Menu Maps Name Editor Menu 204 Name Editor Menu ABCDEFG HIJKLMN Return OPQRSTU VWXYZ Name ...
Page 214: ...Menu Maps SPAN Menu 206 SPAN Menu Span Span Full Span Zero Span Last Span ...
Page 216: ...Menu Maps Trace Det Menu 208 ...
Page 219: ......