133
Key Reference
Meas
Abs AND Rel
Set the limit test to show Fail if one offset’s spectrum
emission mask measurement results is larger than
both the corresponding limit for
Abs Start
and
Abs Stop
and the limit for
Rel Start
and
Rel Stop
.
Abs OR Rel
Set the limit test to show Fail if one offset’s the
absolute spectrum emission mask measurement
results is larger than the corresponding limit for
Abs
Start
and
Abs Stop
or the limit for
Rel Start
and
Rel Stop
.
Meas BW
Specifies the bandwidth to use when measuring the offset. When
using the front panel, only the currently selected offset is affected.
All the offsets are affected when remote control. This is the
multiplier applied to the resolution bandwidth to determine the
bandwidth to integrate when calculating the power. Meas BW is
restricted to be 1 and not changeable.
Key access:
Meas > SEM > Offset/Limits > More > More
Summary of Contents for N9321C
Page 36: ...Getting Started Helpful Tips 28 ...
Page 110: ...Functions and Measurements Files Operation 102 ...
Page 180: ...Key Reference Trace Det 172 ...
Page 212: ...Menu Maps Name Editor Menu 204 Name Editor Menu ABCDEFG HIJKLMN Return OPQRSTU VWXYZ Name ...
Page 214: ...Menu Maps SPAN Menu 206 SPAN Menu Span Span Full Span Zero Span Last Span ...
Page 216: ...Menu Maps Trace Det Menu 208 ...
Page 219: ......