Functions and Measurements
Measuring Multiple Signals
42
Figure 3-7 Unresolved small signal from large signal
5
Reduce the resolution bandwidth filter to view the smaller signal.
The smaller signal will be hidden by the larger signal when the
bandwidth settings are wider. Reducing the RBW setting will allow
less of the larger signal to pass through the analyzer and the
smaller signals peak will then rise out of the noise floor. Place a
delta marker on the smaller signal:
Press
BW > 1 > kHz
Press
Peak Search
Press
Marker > Mode > Delta