Instrument Messages
Device-Specific Errors
180
224
Option not licensed
The selected option requires a license. Refer to the installation
procedures in the user guide provided for this particular option.
238
VBW filter or Average Detector cannot use both
501
RF over load
Either increase input attenuation or decrease the input signal level.
502
IF over load
Either increase input attenuation or decrease the input signal level.
604
File already exists
You attempted to save to a file that already exists. Either delete or
rename the existing file or select a new name, then try again.
605
Media is protected
A save was attempted to a write-protected device.
610
File access is denied
The file is protected or hidden and cannot be accessed.
612
File not found
The analyzer could not find the specified file.
651
Connect TG OUT to RF IN
Attempt to align the tracking generator without its output
connected. Connect the
TG SOURCE
to
RF IN
.
652
Connect CAL OUT to RF IN
you must connect the
CAL OUT
to
RF IN
with the appropriate cable.
751
Instrument state set to initial values
While trying to load a trace or state, the state information was
found to be in error. This may be because the state had been
stored on a later revision of analyzer firmware. A default set of state
variables was loaded instead. There is nothing wrong with the
analyzer.
Summary of Contents for N9321C
Page 36: ...Getting Started Helpful Tips 28 ...
Page 110: ...Functions and Measurements Files Operation 102 ...
Page 180: ...Key Reference Trace Det 172 ...
Page 212: ...Menu Maps Name Editor Menu 204 Name Editor Menu ABCDEFG HIJKLMN Return OPQRSTU VWXYZ Name ...
Page 214: ...Menu Maps SPAN Menu 206 SPAN Menu Span Span Full Span Zero Span Last Span ...
Page 216: ...Menu Maps Trace Det Menu 208 ...
Page 219: ......