Definition of Terms
Keysight
N4373D Lightwave Component Analyzer, User’s Guide
137
Average output power uncertainty (optical)
The maximum difference between actual (true)
and the nominal (selected) average output
power.
Conditions:
After modulator auto bias (“modulator
optimization”) performed with the selected wavelength. Other
conditions as specified.
Measurement:
Using an average power meter with averaging
time set to 1 s.
Average power measurement uncertainty (optical)
Maximum difference between measured average (over time)
optical power and true average optical power at the LCA optical
input.
Measurement:
Comparison with an average power meter with
averaging time set to 1 s.
Average power measurement range (optical)
The range for the average optical power where the specification
for
Average power measurement uncertainty (optical)
applies.
Decibel (dBm, dBo, dBe)
A ratio in decibel (dB) is calculated as 10-log
10
{ratio}. Special
cases:
• dBo: specifically the ratio of optical powers (‘o’ for ‘optical’).
• dBe: specifically the ratio of electrical powers (‘e’ for
‘electrical’).
• dBm: power level related to 1 mW (electrical or optical).
N O T E
Differences of powers in dBm are written as “dBo” for optical powers or
“dBe” for electrical powers.
Summary of Contents for N4373D
Page 1: ...Keysight N4373D Lightwave Component Analyzer User s Guide ...
Page 2: ......
Page 3: ......
Page 38: ......
Page 84: ...User Interface Reference 82 Keysight N4373D Lightwave Component Analyzer User s Guide ...
Page 126: ......
Page 130: ......
Page 136: ......
Page 149: ...Specifications Keysight N4373D Lightwave Component Analyzer User s Guide 147 ...
Page 158: ...Specifications 156 Keysight N4373D Lightwave Component Analyzer User s Guide ...
Page 168: ......
Page 170: ...Ordering Information 168 Keysight N4373D Lightwave Component Analyzer User s Guide ...
Page 172: ......
Page 175: ......