User Interface Reference
74
Keysight
N4373D Lightwave Component Analyzer, User’s Guide
The Measurement Parameters
Measure Mode
A 2-port network analyser can only support single-ended
measurements. Selecting the Measure Mode is only possible if
you have a 4-port network analyzer (applies to 4-port PNA only).
Differential
If you select this mode, means the LCA uses a differential
electrical input or output for this measurement.
Whether the differential signal is the input to the DUT or the
output from the DUT, this uses ports 2 and 3 of a four port
network analyzer.
Single-Ended
If you select this mode, the LCA uses a single-ended electrical
input or output for this measurement.
The electrical input to the DUT is Port A of the optical test set.
The electrical output from the DUT is port B of the optical test
set.
LCA System Settings
Wavelength (nm)
• If you have not selected to use the optional External Input,
select the wavelength of the laser source and the optical
power meter use for measurements.
Summary of Contents for N4373D
Page 1: ...Keysight N4373D Lightwave Component Analyzer User s Guide ...
Page 2: ......
Page 3: ......
Page 38: ......
Page 84: ...User Interface Reference 82 Keysight N4373D Lightwave Component Analyzer User s Guide ...
Page 126: ......
Page 130: ......
Page 136: ......
Page 149: ...Specifications Keysight N4373D Lightwave Component Analyzer User s Guide 147 ...
Page 158: ...Specifications 156 Keysight N4373D Lightwave Component Analyzer User s Guide ...
Page 168: ......
Page 170: ...Ordering Information 168 Keysight N4373D Lightwave Component Analyzer User s Guide ...
Page 172: ......
Page 175: ......