User Interface Reference
Keysight
N4373D Lightwave Component Analyzer, User’s Guide
69
N O T E
After completing the “LCA Measurement Setup” step, the network
analyzer is ready for DUT measurements.
Because the receiver responsivity depends on the optical power, you
should start the DUT measurements with the LCA “Single” or
“Continuous” macros.
These macros perform an optical power measurement prior to each
measurement sweep. This achieves the highest accuracy in EO and OO
measurements.
If you do not need this accuracy, you can also start the DUT measurements
with the network analyzer's trigger functions.
In some cases it can become difficult to abort continuous measurements
by pressing the "Continuous" or "Single" button, since the network
analyzer does not react to the keypad during measurements. If the
"Sweeping..." window (with the "Abort" button) is not shown, abort the
measurement with the "Abort Measurement" button in the LCA Server
window.
For more information on triggering measurements, see and to
.
Summary of Contents for N4373D
Page 1: ...Keysight N4373D Lightwave Component Analyzer User s Guide ...
Page 2: ......
Page 3: ......
Page 38: ......
Page 84: ...User Interface Reference 82 Keysight N4373D Lightwave Component Analyzer User s Guide ...
Page 126: ......
Page 130: ......
Page 136: ......
Page 149: ...Specifications Keysight N4373D Lightwave Component Analyzer User s Guide 147 ...
Page 158: ...Specifications 156 Keysight N4373D Lightwave Component Analyzer User s Guide ...
Page 168: ......
Page 170: ...Ordering Information 168 Keysight N4373D Lightwave Component Analyzer User s Guide ...
Page 172: ......
Page 175: ......