Making Measurements
Keysight
N4373D Lightwave Component Analyzer, User’s Guide
89
Measuring Opto-Electrical (OE) Devices
Always perform an electrical calibration of the network
analyzer or load an existing calibration set before you make a
measurement.
For standard measurements, see “Electrical calibration for
single-ended measurements” on page 85.
For differential measurements (applies to 4-port PNA only), see
“Electrical Calibration for Differential OE Measurements” on
page 86.
For OE measurements the single-ended (SE), logical port 1 is
assigned to physical port 1 and the balanced (BAL), logical port
2 is assigned to the physical ports 2 and 3.
For OE measurements, the LCA sets up traces for S22, S21, S33,
S31 by default.
For more information, see “Differential Measurements (applies
to 4-port PNA only)” on page 59.
Before you make your first opto-electrical measurement, you
must initialise the OE measurement. You can see which
measurement is currently initialised at the top left of the trace
screen.
If you have already initialised the OE measurement, you can
connect your device under test and make the measurement by
pressing
or
.
Please refer to the user guide and online help for the network
analyzer for further information.
Summary of Contents for N4373D
Page 1: ...Keysight N4373D Lightwave Component Analyzer User s Guide ...
Page 2: ......
Page 3: ......
Page 38: ......
Page 84: ...User Interface Reference 82 Keysight N4373D Lightwave Component Analyzer User s Guide ...
Page 126: ......
Page 130: ......
Page 136: ......
Page 149: ...Specifications Keysight N4373D Lightwave Component Analyzer User s Guide 147 ...
Page 158: ...Specifications 156 Keysight N4373D Lightwave Component Analyzer User s Guide ...
Page 168: ......
Page 170: ...Ordering Information 168 Keysight N4373D Lightwave Component Analyzer User s Guide ...
Page 172: ......
Page 175: ......