![Keysight Technologies N4373D User Manual Download Page 105](http://html.mh-extra.com/html/keysight-technologies/n4373d/n4373d_user-manual_1962029105.webp)
Making Measurements
Keysight
N4373D Lightwave Component Analyzer, User’s Guide
103
If you have already characterized the device and have a .s2p
file, for “Deembedding Data”, select “File”, and enter or select
the file.
• For accurate measurements, you can also include the
optical path in the test circuit without the DUT, and select
“Measure” (or “Load”) for User Calibration.
8 Decide what sort of calibration you need.
• If for “User Calibration”, you select “None”, the LCA uses
the factory default calibration.
• The factory calibration ensures the uncertainty specified,
but particularly for improved results at lower frequencies,
we recommend an extra calibration. In this case, for “User
Calibration”, select “Measure”.
• If you want to reuse and existing, saved calibration, for
“User Calibration”, select “Load”, and select the file from
the Windows Explorer window.
9 Click
to start the measurement preparation.
If you are calibrating, when the following screen is displayed,
make sure the optical output is connected to the optical input
using the patchcord supplied.
The patchcord has the following part number:
• Single-mode angled/angled connector: N4373-87906
• Single-mode angled/straight connector: N4373-87907
N O T E
Optical path and electrical path deembedding is not applied during a user
calibration.
Summary of Contents for N4373D
Page 1: ...Keysight N4373D Lightwave Component Analyzer User s Guide ...
Page 2: ......
Page 3: ......
Page 38: ......
Page 84: ...User Interface Reference 82 Keysight N4373D Lightwave Component Analyzer User s Guide ...
Page 126: ......
Page 130: ......
Page 136: ......
Page 149: ...Specifications Keysight N4373D Lightwave Component Analyzer User s Guide 147 ...
Page 158: ...Specifications 156 Keysight N4373D Lightwave Component Analyzer User s Guide ...
Page 168: ......
Page 170: ...Ordering Information 168 Keysight N4373D Lightwave Component Analyzer User s Guide ...
Page 172: ......
Page 175: ......