Making Measurements
Keysight
N4373D Lightwave Component Analyzer, User’s Guide
93
• Normally, you do not need extra calibration, that is, select
“None”. The LCA measures to specifications under the
standard conditions.
• The frequency dependent transmission behavior of optical
transmitters cannot be characterized as accurately as the
behavior of receivers. The factory calibration ensures the
uncertainty specified, but particularly for improved
results at lower frequencies, this may be slightly improved
by performing a user calibration. In this case, select
“Measure”.
• If you want to reuse an existing, saved calibration, select
“Load”, and select the file from the Windows Explorer
window.
10 Click
to start the measurement preparation.
If you are performing a user calibration , when the following
screen is displayed, make sure the optical output is
connected to the optical input using the patchcord supplied.
The patchcord has the following part number:
• Single-mode angled/angled connector: N4373-87906
• Single-mode angled/straight connector: N4373-87907
Please also refer to the documentation supplied with the
patchcord.
a Select “Reduce IFBW for user calibration measurement” to
reduce noise on the user calibration data. The default
N O T E
Optical path and electrical path deembedding is not applied during a user
calibration.
Summary of Contents for N4373D
Page 1: ...Keysight N4373D Lightwave Component Analyzer User s Guide ...
Page 2: ......
Page 3: ......
Page 38: ......
Page 84: ...User Interface Reference 82 Keysight N4373D Lightwave Component Analyzer User s Guide ...
Page 126: ......
Page 130: ......
Page 136: ......
Page 149: ...Specifications Keysight N4373D Lightwave Component Analyzer User s Guide 147 ...
Page 158: ...Specifications 156 Keysight N4373D Lightwave Component Analyzer User s Guide ...
Page 168: ......
Page 170: ...Ordering Information 168 Keysight N4373D Lightwave Component Analyzer User s Guide ...
Page 172: ......
Page 175: ......