![Keysight Technologies N4373D User Manual Download Page 48](http://html.mh-extra.com/html/keysight-technologies/n4373d/n4373d_user-manual_1962029048.webp)
Performance Quick Check
46
Keysight
N4373D Lightwave Component Analyzer, User’s Guide
7. Select the OO measurement tab.
8. Set the following parameters:
• If you have option 102 (Dual wavelength 1310nm,
1550nm), set Wavelength to the value at which you are
testing your DUTs (1310nm or 1550nm).
• Laser power: +4 dBm
• Optical input: input1 (“+7 dBm max”) or input 2
(“+17 dBm max”)
• User Calibration: None
• Forward RF Power: -1 dBm
9. Click
, then
.
10.Right click the measurement trace, and select “Autoscale All”
in the context menu.
At the end of the measurement, there should be a flat trace.