Measurement Concepts
60
Keysight
N4373D Lightwave Component Analyzer, User’s Guide
Differential OE measurements
For OE measurements the single-ended (SE), logical port 1 is
assigned to physical port 1 and the balanced (BAL), logical port
2 is assigned to the physical ports 2 and 3 (see Figure 4).
Figure 4
Logical Port Mapping for OE measurement
For OE measurements, the LCA sets up traces for S22, S21, S33,
S31 by default.
Differential EO measurements
For EO measurement the single-ended (SE), logical port 1 is
assigned to physical port 4 and the balanced (BAL), logical port
2 is assigned to physical ports 2 and 3 (see Figure 5)
Figure 5
Logical Port Mapping for EO measurement
For EO measurements the LCA sets up traces for S22, S42, S33,
S43 by default.