Testing Performance
2
6000L Series Oscilloscopes Service Guide
25
To verify digital channel threshold accuracy (MSO models only)
This test verifies the digital channel threshold accuracy specification of the
Keysight 6000L Series Oscilloscopes.
Threshold accuracy test limits= ±(100 mV + 3% of threshold setting)
When to Test
You should perform this test every 12 months or after 2000 hours of operation,
whichever comes first.
What to Test
Use these instructions to test the threshold settings of digital channels D7-D0.
Then, use the same instructions to test digital channels D15-D8.
Verifying Test Results
After each threshold test, record the voltage reading in the Performance Test
. To verify whether a test passes, verify that the voltage reading
is within the limits in the Performance Test Record.
Table 4
Equipment Required to Test Digital Channel Threshold Accuracy
Equipment
Critical Specifications
Recommended Model/Part
Digital Multimeter
0.1 mV resolution,
0.005% accuracy
Keysight 34401A
Oscilloscope Calibrator
DC offset voltage 6.3 V Fluke 5820A
BNC-Banana Cable
Keysight 11001-60001
BNC Tee
Keysight 1250-0781
BNC Cable
Fluke 50
Ω
cable, P/N 686318
BNC Test Connector, 8-by-2
Probe Cable
Keysight 01650-61607
Summary of Contents for InfiniVision 6000L Series
Page 1: ...Keysight InfiniVision 6000L Series Oscilloscopes Service Guide...
Page 8: ...8 6000L Series Oscilloscopes Service Guide Contents...
Page 10: ...10 6000L Series Oscilloscopes Service Guide Figures...
Page 12: ...12 6000L Series Oscilloscopes Service Guide Tables...
Page 48: ...48 6000L Series Oscilloscopes Service Guide 2 Testing Performance...
Page 54: ...54 6000L Series Oscilloscopes Service Guide 3 Calibrating and Adjusting...
Page 86: ...86 6000L Series Oscilloscopes Service Guide 5 Replacing 6000L Assemblies...
Page 94: ...94 6000L Series Oscilloscopes Service Guide 6 Replaceable Parts...