16
6000L Series Oscilloscopes Service Guide
1
Characteristics and Specifications
6000L Series Maximum Input Voltages
CAUTION
Maximum input voltage at analog inputs
Do not exceed the maximum input voltage rating.
The maximum input voltage for the 50
Ω
input impedance setting is ±5 V.
Input protection is enabled in 50
Ω
mode and the 50
Ω
load will disconnect if greater
than 5 Vrms is detected. However the inputs could still be damaged, depending on the
time constant of the signal. The 50
Ω
input protection functions only when the
oscilloscope is powered on.
The maximum input voltage for the 1 M
Ω
input impedance setting is 30 Vrms or
±40 Vmax (DC+Vpeak)
Probing technology allows for testing of higher voltages; the included 10073D 10:1 probe
supports 300 Vrms or ±400 Vmax (DC+Vpeak). No transient overvoltage allowed.
CAUTION
Maximum input voltage at digital channels
±40 V peak