66
6000L Series Oscilloscopes Service Guide
4
Troubleshooting
Figure 11
System Board Test Points
5
If the voltage is not correct:
a
Disconnect the power cord.
b
Remove the fan. See
c
Disconnect plug J3 from the power supply and from the system board.
d
Perform a continuity check on the “DC cable” (item
on
). If the
cable is shorted or open, replace it. If not, continue to the next step.
e
Reconnect the “DC cable” to the power supply, but not to the system board.
Pin 1
Pin 9
J3
200