Testing Performance
2
6000L Series Oscilloscopes Service Guide
21
List of Test Equipment
Below is a list of test equipment and accessories required to perform the
performance test verification procedures.
Table 2
List of test equipment
Equipment
Critical Specifications
Recommended Model/
Part Number
Test connector, 8-by-2*
for instructions on building test connector.
n/a
Digital Multimeter
0.1 mV resolution, 0.005% accuracy
Keysight 34401A
Power Splitter
Outputs differ by 0.15 dB
Keysight 11667B
Oscilloscope Calibrator
DC offset voltage of -5.5 V to 35.5 V, 0.1 V resolution
Fluke 5820A
Signal Generator
25 MHz, 100 MHz, 500 MHz, and 1 GHz sine waves
Keysight N5181A, or
E4400B, or 8648A
Power Meter/Sensor
1 GHz
±
3% accuracy
Keysight E4418B/8482A
Oscilloscope Calibrator
25 MHz—500 MHz sine wave, 5 ppm
Fluke 5820A
BNC banana cable
Keysight 11001-66001
BNC cable (qty 3)
Keysight 10503A
Cable
Type N (m) 609.6 mm (24 in.)
Keysight 11500B
Probe cable*
Keysight 01650-61607
Shorting Cap BNC
Keysight 1250-0774
Adapter
BNC(f) to banana(m)
Keysight 1251-2277
Adapter
BNC Tee (m) (f) (f)
Keysight 1250-0781
Adapter
Type N (m) to BNC (m)
Keysight 1250-0082
or Pomona 3288 with
Pomona 3533
Blocking capacitor
Keysight 10240-60001
Adapter (qty 3)
N(m) to BNC(f)
Keysight 1250-0780
Feedthrough (qty 2)
50
Ω
BNC (f) to BNC (m)
Keysight 0960-0301
* Required only for testing digital channels of oscilloscopes that have the MSO option.
Some parts and equipment are available at www.keysight.com or www.parts.keysight.com.
Summary of Contents for InfiniVision 6000L Series
Page 1: ...Keysight InfiniVision 6000L Series Oscilloscopes Service Guide...
Page 8: ...8 6000L Series Oscilloscopes Service Guide Contents...
Page 10: ...10 6000L Series Oscilloscopes Service Guide Figures...
Page 12: ...12 6000L Series Oscilloscopes Service Guide Tables...
Page 48: ...48 6000L Series Oscilloscopes Service Guide 2 Testing Performance...
Page 54: ...54 6000L Series Oscilloscopes Service Guide 3 Calibrating and Adjusting...
Page 86: ...86 6000L Series Oscilloscopes Service Guide 5 Replacing 6000L Assemblies...
Page 94: ...94 6000L Series Oscilloscopes Service Guide 6 Replaceable Parts...