Testing Performance
2
6000L Series Oscilloscopes Service Guide
47
Horizontal
Δ
t Accuracy
Generator Setting
Test Limits
Results
Period
100
μ
s
99.8
μ
s to 100.2
μ
s
________
Period
100 ns
99.8 ns to 100.2 ns
________
Period
10 ns
1
9.93 ns to 10.07 ns
________
Period
5 ns
2
4.96 ns to 5.04 ns
________
1
100 MHz models
2
500 MHz and 1 GHz models
Trigger Sensitivity
Test Limits
Channel 1
Channel 2
Channel 3*
Channel 4*
Internal trigger
1 division at 25 MHz
________
________
________
________
1 GHz models
0.6 division at 1 GHz
________
________
________
_______
500 MHz models
0.6 division at 500 MHz ________
________
________
_______
100 MHz models
0.6 division at 100 MHz ________
________
________
_______
Trigger Sensitivity
Test Limits
Channel 1
Channel 2
Channel 3*
Channel 4*
External trigger (6014, 6054, 6104)
6104, 6054
500 mV at 500 MHz
________
________
________
________
6014
500 mV at 100 MHz
________
________
________
________
* Where applicable
Summary of Contents for InfiniVision 6000L Series
Page 1: ...Keysight InfiniVision 6000L Series Oscilloscopes Service Guide...
Page 8: ...8 6000L Series Oscilloscopes Service Guide Contents...
Page 10: ...10 6000L Series Oscilloscopes Service Guide Figures...
Page 12: ...12 6000L Series Oscilloscopes Service Guide Tables...
Page 48: ...48 6000L Series Oscilloscopes Service Guide 2 Testing Performance...
Page 54: ...54 6000L Series Oscilloscopes Service Guide 3 Calibrating and Adjusting...
Page 86: ...86 6000L Series Oscilloscopes Service Guide 5 Replacing 6000L Assemblies...
Page 94: ...94 6000L Series Oscilloscopes Service Guide 6 Replaceable Parts...