To remove the AC power input assembly
1
Perform the previous procedures:
•
•
•
2
Using a TORX T10 screwdriver, remove the screw securing the power
receptacle ground wire to the chassis.
3
Using a TORX T10 screwdriver, remove the screw securing the input board to
the chassis.
4
Lift and rotate the grooved input receptacle just enough to free it from the
opening in the chassis.
5
Slide the assembly toward the front of the chassis to remove the posts from the
keyholes.
6
Lift the whole assembly up and back to free the power shaft from the chassis.
CAUTION
Avoid breaking the power shaft. Use one hand on each of the 2 pieces of the assembly to
keep it in alignment as you remove it.