Replacing 6000L Assemblies
5
6000L Series Oscilloscopes Service Guide
79
To remove the power supply
1
Perform the previous procedures:
•
•
2
Disconnect the ground cable and AC cables from the power supply using a
needle nose pliers if necessary.
3
Press the release latch to disconnect the power input board cable from the
power input board.
Figure 18
Removing the power supply.
4
Using a T10 Torx driver, remove the 2 screws securing the power supply to the
chassis.
Power Input Board
Cable
T10 Screws
Power Input Board
Ground
Connector
Hole in
Divider
AC
Connector
Snap
Fasteners
Power to System
Board