6000L Series Oscilloscopes Service Guide
2
Testing Performance
Test Internal Trigger Sensitivity
1
Connect the equipment (see
a
For 500 MHz and 1 GHz models, connect the signal generator output to the
oscilloscope channel 1 input.
b
For 100 MHz models, use a 50
Ω
feedthrough termination to connect the
signal generator output to the oscilloscope channel 1 input.
Figure 6
Connect equipment for internal trigger sensitivity test
Signal
Generator
N to BNC adapter
N Cable
50
Ω
Feedthrough
*
*
100 MHz models only.
Oscilloscope