Testing Performance
2
6000L Series Oscilloscopes Service Guide
39
To verify horizontal delta t accuracy
The following test is optional because horizontal
Δ
t accuracy is a characteristic,
not a specification.
This test verifies the horizontal
Δ
t accuracy. In this test, you will use the
oscilloscope to measure the output of a time mark generator.
Test limits: ±0.0015% of reading ±0.1% of full scale ±20 ps (same channel)
1
Connect the equipment:
a
For 500 MHz and 1 GHz oscilloscope models, connect the calibrator output
to the oscilloscope channel 1 input.
b
For 100 MHz oscilloscopes, use a 50
Ω
feedthrough termination to connect
the calibrator output to the oscilloscope channel 1 input.
2
Set up the signal source.
a
Select
Marker
on the oscilloscope calibrator.
b
Set the calibrator for 100
μ
s markers (period = 100
μ
s).
3
Set up the oscilloscope.
a
Set channel 1
Coupling
to
DC
.
b
On 1 GHz and 500 MHz models only, set channel 1
Imped
to
50 Ohm
.
c
Press the
Display
key,
then set the
Vectors
softkey to off.
d
Press the
AutoScale
key.
e
Set the time base to 20
μ
s/div.
f
Press the
Main/Delayed
(or
Zoom
) key, then set the
Time Ref
softkey to
Left
.
g
Adjust the Trigger Level knob to obtain a stable display.
Table 10
Equipment Required to Verify Horizontal
Δ
t Accuracy
Equipment
Critical Specifications
Recommended Model/Part
Oscilloscope Calibrator Stability 5 ppm after 1/2 hour
Fluke 5820A
Cable
BNC, 3 feet
Keysight 10503A
Feedthrough
50
Ω
BNC (f) to BNC (m)
Keysight 0960-0301
Summary of Contents for InfiniVision 6000L Series
Page 1: ...Keysight InfiniVision 6000L Series Oscilloscopes Service Guide...
Page 8: ...8 6000L Series Oscilloscopes Service Guide Contents...
Page 10: ...10 6000L Series Oscilloscopes Service Guide Figures...
Page 12: ...12 6000L Series Oscilloscopes Service Guide Tables...
Page 48: ...48 6000L Series Oscilloscopes Service Guide 2 Testing Performance...
Page 54: ...54 6000L Series Oscilloscopes Service Guide 3 Calibrating and Adjusting...
Page 86: ...86 6000L Series Oscilloscopes Service Guide 5 Replacing 6000L Assemblies...
Page 94: ...94 6000L Series Oscilloscopes Service Guide 6 Replaceable Parts...