Testing Performance
2
6000L Series Oscilloscopes Service Guide
45
4
Set the power meter Cal Factor % to the appropriate value (500 MHz or
100 MHz) on the calibration chart on the power sensor. If necessary, do a linear
interpolation if a 500 MHz or 100 MHz factor is not included in the power
meter’s calibration chart.
5
Adjust the signal generator output for reading on the power meter of 625
μ
W.
(500mV
pp
= 176.78mV rms,
Power = Vin
2
/50
Ω
= 176.78 mV
2
/50
Ω
= 625
μ
W.)
6
Press the Trigger
Edge
key, then press the
Source
softkey to set the trigger
source to
External
.
7
Check for stable triggering and adjust the trigger level if necessary. Triggering
is indicated by the
Trig’d
indicator at the top of the display. When it is flashing,
the oscilloscope is not triggered. When it is not flashing, the oscilloscope is
triggered.
8
Record the results as Pass or Fail in the Performance Test Record (see
If the test fails, see the “Troubleshooting” chapter. Then return here.
Summary of Contents for InfiniVision 6000L Series
Page 1: ...Keysight InfiniVision 6000L Series Oscilloscopes Service Guide...
Page 8: ...8 6000L Series Oscilloscopes Service Guide Contents...
Page 10: ...10 6000L Series Oscilloscopes Service Guide Figures...
Page 12: ...12 6000L Series Oscilloscopes Service Guide Tables...
Page 48: ...48 6000L Series Oscilloscopes Service Guide 2 Testing Performance...
Page 54: ...54 6000L Series Oscilloscopes Service Guide 3 Calibrating and Adjusting...
Page 86: ...86 6000L Series Oscilloscopes Service Guide 5 Replacing 6000L Assemblies...
Page 94: ...94 6000L Series Oscilloscopes Service Guide 6 Replaceable Parts...