46
6000L Series Oscilloscopes Service Guide
2
Testing Performance
Keysight 6000L Series Oscilloscopes Performance Test Record
Serial No. ______________________________________
Test by _____________________________
Test Interval ____________________________________
Work Order No. ______________________
Recommended Next Testing ________________________
Temperature ____________
Threshold
Specification
Limits
Ch D7-D0
Ch D15-D8
Accuracy Test
5 V - 250 mV
4.750 V
________
________
(100 mV + 3% of
threshold setting)
5 V + 250 mV
5.250 V
________
________
-5 V - 250 mV
-5.250 V
________
________
-5 V + 250 mV
-4.750 V
________
________
0 V - 100 mV
-100 mV
________
________
0 V + 100 mV
100 mV
________
________
Voltage Measurement Accuracy
Range
Power Supply Setting Test Limits
Channel 1
Channel 2
Channel 3*
Channel 4*
5 V/Div
35 V
34.04 V to 35.96 V
________
________
________
________
2 V/Div
14 V
13.616 V to 14.384 V
________
________
________
________
1 V/Div
7 V
6.808 V to 7.192 V
________
________
________
________
500 mV/Div
3.5 V
3.404 V to 3.596 V
________
________
________
________
200 mV/Div
1.4 V
1.3616 V to 1.4384 V
________
________
________
________
100 mV/Div
700 mV
680.8 mV to 719.2 mV ________
________
________
________
50 mV/Div
350 mV
340.4 mV to 359.6 mV ________
________
________
________
20 mV/Div
140 mV
136.16 mV to 143.84 mV ________
________
________
________
10 mV/Div
70 mV
68.08 mV to 71.92 mV ________
________
________
________
5 mV/Div
35 mV
34.04 mV to 35.96 mV ________
________
________
________
2 mV/Div
(500 MHz and 1 GHz models)
14 mV
13.232 mV to 14.768 mV ________
________
________
________
2 mV/Div
(100 MHz models)
14 mV
13.616 mV to 14.384 mV ________
________
________
________
1 mV/Div
(MSO/DSO6012A)
7 mV
6.616 mV to 7.384 mV ________
________
________
________
Bandwidth
Model
Test Limits
Channel 1
Channel 2
Channel 3*
Channel 4*
610x
-3 dB at 1 GHz
________
________
________
________
605x
-3 dB at 500 MHz
________
________
________
________
601x
-3 dB at 100 MHz
________
________
________
________
Time Scale
Accuracy
Limits
Time Scale Accuracy is a warranted specification
for 6000L Series instruments manufactured after
January 1, 2008.
Calculated
TSA Limit
(ppm)
Measured
time scale
error (ppm)
Pass/Fail
TSA Limit: ± (15 + 2 * YrsSinceMfr) ppm
___________
___________
___________
Summary of Contents for InfiniVision 6000L Series
Page 1: ...Keysight InfiniVision 6000L Series Oscilloscopes Service Guide...
Page 8: ...8 6000L Series Oscilloscopes Service Guide Contents...
Page 10: ...10 6000L Series Oscilloscopes Service Guide Figures...
Page 12: ...12 6000L Series Oscilloscopes Service Guide Tables...
Page 48: ...48 6000L Series Oscilloscopes Service Guide 2 Testing Performance...
Page 54: ...54 6000L Series Oscilloscopes Service Guide 3 Calibrating and Adjusting...
Page 86: ...86 6000L Series Oscilloscopes Service Guide 5 Replacing 6000L Assemblies...
Page 94: ...94 6000L Series Oscilloscopes Service Guide 6 Replaceable Parts...