2-18
Voltage and Temperature Measurements
Temperature configuration
If you are going to perform temperature measurements, you have to configure the Model 2182
appropriately from the temperature configuration menu:
Temperature configuration menu
The items of the temperature configuration menu are explained as follows:
•
UNITS
— Select the desired units designator for temperature readings (˚C, ˚F, or K).
•
SENS
— Select the thermocouple (TCOUPLE) to perform temperature measurements
at the thermocouple. The internal (INTERNL) sensor is used to measure the internal
temperature of the Model 2182.
•
TYPE
— Select the thermocouple type that you are using to measure temperature (J, K,
T, E, R, S, B, or N).
•
JUNC
— Select INTRNL to reference measurements to the internal reference junction.
Select SIM to reference measurements to an external simulated reference. After selecting
SIM, you will be prompted to enter the simulated reference temperature.
After pressing SHIFT and then TCOUP to access the menu, use the following rules to
configure temperature:
•
There are four menu items; UNITS, SENS, TYPE, and JUNC. Along with each menu
item, the present option is displayed. For example, if ˚C is the present units option, then
“UNITS: C” is displayed.
•
Blinking characters indicate cursor position. The cursor can be on a menu item name (i.e.
“UNITS” blinking) or on an menu item option (i.e. “C” blinking). Cursor position is
controlled by the
and
keys.
•
With the cursor on a menu item name, you can use the
or
key to scroll through the
other menu items. Pressing ENTER will select the displayed option and move on to the
next menu item (or exit if at end of menu).
•
With the cursor on a menu item option, you can use the
or
key to display one of
the other options for that menu item. Pressing ENTER will select the displayed option
and move on to the next menu item (or exit if at end of menu). An exception is the SIM
menu item. After selecting SIM, you will be prompted to enter the simulated
temperature. Use the arrow keys to display the value and press ENTER.
•
Pressing EXIT leaves the menu and returns to the normal display state.
Summary of Contents for 2182
Page 1: ...www tek com keithley Model 2182 2182A Nanovoltmeter User s Manual 2182A 900 01 Rev B May 2017...
Page 18: ......
Page 22: ......
Page 23: ...1 Getting Started Getting Started...
Page 41: ...2 VoltageandTemperature Measurements Voltageand Temperature Measurements...
Page 68: ...2 28 Voltage and Temperature Measurements...
Page 69: ...3 Range Digits Rate andFilter Range Digits Rate andFilter...
Page 82: ...3 14 Range Digits Rate and Filter...
Page 83: ...4 Relative mX b and Percent Relative mX b andPercent...
Page 91: ...5 RatioandDelta Ratioand Delta...
Page 117: ...6 Buffer Buffer...
Page 123: ...7 Triggering Triggering...
Page 140: ...7 18 Triggering...
Page 141: ...8 Limits Limits...
Page 149: ...9 SteppingandScanning Steppingand Scanning...
Page 168: ...9 20 Stepping and Scanning...
Page 169: ...10 AnalogOutput Analog Output...
Page 175: ...11 RemoteOperation Remote Operation...
Page 205: ...12 CommonCommands Common Commands...
Page 221: ...13 SCPISignalOriented Measurement Commands SCPISignalOri entedMeasure ment Commands...
Page 225: ...14 SCPIReferenceTables SCPIRefer enceTables...
Page 239: ...15 AdditionalSCPI Commands Additional SCPICom mands...
Page 260: ...15 22 Additional SCPI Commands...
Page 261: ...A Specifications Specifications...
Page 263: ...B Statusand ErrorMessages StatusandError Messages...
Page 268: ...B 6 Status and Error Messages...
Page 269: ...C Measurement Considerations Measurement Consider ations...
Page 278: ...C 10 Measurement Considerations...
Page 279: ...D Model182Emulation Commands Model182 EmulationCom mands...
Page 284: ...D 6 Model 182 Emulation Commands...
Page 285: ...E Example Programs ExamplePro grams...
Page 293: ...F IEEE 488 BusOverview IEEE 488Bus Overview...
Page 307: ...G IEEE 488andSCPI ConformanceInformation IEEE 488and SCPIConform anceInforma tion...
Page 310: ...G 4 IEEE 488 and SCPI Conformance Information...
Page 311: ...H Measurement Queries Measurement Queries...
Page 316: ...H 6 Measurement Queries...
Page 317: ...I Delta PulseDelta and DifferentialConductance Delta Pulse Deltaand Dif ferentialCon ductance...