Remote Operation
11-15
Condition registers
As
shows, some register sets have a condition register. A condition register is a
real-time, read-only register that constantly updates to reflect the present operating conditions
of the instrument. For example, while a measurement is being performed, bit B4 (Meas) of the
Operation Condition Register is set. When the measurement is completed, bit B4 clears.
Use the :CONDition? query commands in the STATus Subsystem to read the condition
registers. See
Event registers
As
shows, each status register set has an event register. An event register is a
latched, read-only register whose bits are set by the corresponding condition register. Once a bit
in an event register is set, it remains set (latched) until the register is cleared by a specific clearing
operation. The bits of an event register are logically ANDed with the bits of the corresponding
enable register and applied to an OR gate. The output of the OR gate is applied to the Status Byte
Register.
Use the *ESR? Common Command to read the Standard Event Register. All other event
registers are read using the EVENt? query commands in the STATus Subsystem. See
for more information.
An event register is cleared when it is read. The following operations clear all event registers:
•
Cycling power
•
Sending *CLS
Enable registers
As
shows, each status register set has an enable register. An enable register is
programmed by you and serves as a mask for the corresponding event register. An event bit is
masked when the corresponding bit in the enable register is cleared (0). When masked, a set bit
in an event register cannot set a bit in the Status Byte Register (1 AND 0 = 0).
To use the Status Byte Register to detect events (i.e., serial poll), you must unmask the events
by setting (1) the appropriate bits of the enable registers.
To program and query the Standard Event Status Register, use the *ESE and *ESE? Common
Commands respectively. All other enable registers are programmed and queried using the
:ENABle and :ENABle? commands in the STATus Subsystem. See
information.
An enable register is not cleared when it is read. The following operations affect the enable
registers:
•
Cycling power — Clears all enable registers
•
:STATus:PRESet clears the following enable registers:
•
Operation Event Enable Register
•
Questionable Event Enable Register
•
Measurement Event Enable Register
•
*ESE 0 — Clears the Standard Event Status Enable Register
Summary of Contents for 2182
Page 1: ...www tek com keithley Model 2182 2182A Nanovoltmeter User s Manual 2182A 900 01 Rev B May 2017...
Page 18: ......
Page 22: ......
Page 23: ...1 Getting Started Getting Started...
Page 41: ...2 VoltageandTemperature Measurements Voltageand Temperature Measurements...
Page 68: ...2 28 Voltage and Temperature Measurements...
Page 69: ...3 Range Digits Rate andFilter Range Digits Rate andFilter...
Page 82: ...3 14 Range Digits Rate and Filter...
Page 83: ...4 Relative mX b and Percent Relative mX b andPercent...
Page 91: ...5 RatioandDelta Ratioand Delta...
Page 117: ...6 Buffer Buffer...
Page 123: ...7 Triggering Triggering...
Page 140: ...7 18 Triggering...
Page 141: ...8 Limits Limits...
Page 149: ...9 SteppingandScanning Steppingand Scanning...
Page 168: ...9 20 Stepping and Scanning...
Page 169: ...10 AnalogOutput Analog Output...
Page 175: ...11 RemoteOperation Remote Operation...
Page 205: ...12 CommonCommands Common Commands...
Page 221: ...13 SCPISignalOriented Measurement Commands SCPISignalOri entedMeasure ment Commands...
Page 225: ...14 SCPIReferenceTables SCPIRefer enceTables...
Page 239: ...15 AdditionalSCPI Commands Additional SCPICom mands...
Page 260: ...15 22 Additional SCPI Commands...
Page 261: ...A Specifications Specifications...
Page 263: ...B Statusand ErrorMessages StatusandError Messages...
Page 268: ...B 6 Status and Error Messages...
Page 269: ...C Measurement Considerations Measurement Consider ations...
Page 278: ...C 10 Measurement Considerations...
Page 279: ...D Model182Emulation Commands Model182 EmulationCom mands...
Page 284: ...D 6 Model 182 Emulation Commands...
Page 285: ...E Example Programs ExamplePro grams...
Page 293: ...F IEEE 488 BusOverview IEEE 488Bus Overview...
Page 307: ...G IEEE 488andSCPI ConformanceInformation IEEE 488and SCPIConform anceInforma tion...
Page 310: ...G 4 IEEE 488 and SCPI Conformance Information...
Page 311: ...H Measurement Queries Measurement Queries...
Page 316: ...H 6 Measurement Queries...
Page 317: ...I Delta PulseDelta and DifferentialConductance Delta Pulse Deltaand Dif ferentialCon ductance...