SCPI Signal Oriented Measurement Commands
13-3
•
Buffer operation is disabled. A storage operation presently in process will be aborted.
•
Autozero is set to the *RST default value.
•
All operations associated with stepping or scanning are disabled.
This command is automatically asserted when the :MEASure? command is sent.
Program fragment
CALL SEND (7, “:conf:volt”, status%)
‘ Perform CONFigure operations.
CALL SEND (7, “:trig:del 0.5”, status%)
‘ Set delay for 0.5sec.
CALL SEND (7, “:samp:coun 10”, status%)
‘ Set sample count to 10.
CALL SEND (7, “:read?”, status%)
‘ Trigger and request readings.
reading$ = SPACE$(80)
CALL ENTER(reading$, length%, 7, status%) ‘ Address 2182 to talk.
PRINT reading$
‘ Display the 10 readings on the
‘ CRT.
:FETCh?
Description
This command requests the latest post-processed reading. After sending this command and
addressing the Model 2182 to talk, the reading is sent to the computer. This command does not
affect the instrument setup.
This command does not trigger a measurement. The command simply requests the last
available reading. Note that this command can repeatedly return the same reading. Until there is
a new reading, this command continues to return the old reading. If your application requires a
“fresh” reading, use the :DATA:FRESh? command (see the SENSe Subsystem command).
This command is automatically asserted when the :READ? or :MEASure? command is sent.
:READ?
Description
This command is typically used with the instrument in the “one-shot” measurement mode to
trigger and acquire a specified number of readings. The :SAMPle:COUNt command is used to
specify the number of readings (see Trigger Subsystem). Note that the readings are stored in the
buffer.
When this command is sent, the following commands execute in the order they are presented:
:ABORt
:INITiate
:FETCh?
When :ABORt is executed, the instrument goes into the idle state if continuous initiation is
disabled. If continuous initiation is enabled, the operation re-starts at the beginning of the
Trigger Model.
Summary of Contents for 2182
Page 1: ...www tek com keithley Model 2182 2182A Nanovoltmeter User s Manual 2182A 900 01 Rev B May 2017...
Page 18: ......
Page 22: ......
Page 23: ...1 Getting Started Getting Started...
Page 41: ...2 VoltageandTemperature Measurements Voltageand Temperature Measurements...
Page 68: ...2 28 Voltage and Temperature Measurements...
Page 69: ...3 Range Digits Rate andFilter Range Digits Rate andFilter...
Page 82: ...3 14 Range Digits Rate and Filter...
Page 83: ...4 Relative mX b and Percent Relative mX b andPercent...
Page 91: ...5 RatioandDelta Ratioand Delta...
Page 117: ...6 Buffer Buffer...
Page 123: ...7 Triggering Triggering...
Page 140: ...7 18 Triggering...
Page 141: ...8 Limits Limits...
Page 149: ...9 SteppingandScanning Steppingand Scanning...
Page 168: ...9 20 Stepping and Scanning...
Page 169: ...10 AnalogOutput Analog Output...
Page 175: ...11 RemoteOperation Remote Operation...
Page 205: ...12 CommonCommands Common Commands...
Page 221: ...13 SCPISignalOriented Measurement Commands SCPISignalOri entedMeasure ment Commands...
Page 225: ...14 SCPIReferenceTables SCPIRefer enceTables...
Page 239: ...15 AdditionalSCPI Commands Additional SCPICom mands...
Page 260: ...15 22 Additional SCPI Commands...
Page 261: ...A Specifications Specifications...
Page 263: ...B Statusand ErrorMessages StatusandError Messages...
Page 268: ...B 6 Status and Error Messages...
Page 269: ...C Measurement Considerations Measurement Consider ations...
Page 278: ...C 10 Measurement Considerations...
Page 279: ...D Model182Emulation Commands Model182 EmulationCom mands...
Page 284: ...D 6 Model 182 Emulation Commands...
Page 285: ...E Example Programs ExamplePro grams...
Page 293: ...F IEEE 488 BusOverview IEEE 488Bus Overview...
Page 307: ...G IEEE 488andSCPI ConformanceInformation IEEE 488and SCPIConform anceInforma tion...
Page 310: ...G 4 IEEE 488 and SCPI Conformance Information...
Page 311: ...H Measurement Queries Measurement Queries...
Page 316: ...H 6 Measurement Queries...
Page 317: ...I Delta PulseDelta and DifferentialConductance Delta Pulse Deltaand Dif ferentialCon ductance...