14-4
SCPI Reference Tables
Command
Description
Default
Parameter
Ref
SCPI
[:IMMediate]
Clear limit test results.
√
:AUTO <b>
Enable or disable clearing of limit test results
when a new trigger model cycle starts.
ON
√
:AUTO?
Query state of auto clear.
√
:LIMit2
Limit 2 Testing:
√
:UPPer
Configure upper limit:
√
[:DATA] <n>
Specify limit; -100e6 to 100e6.
2
√
[:DATA]?
Query upper limit.
√
:LOWer
Configure lower limit:
√
[:DATA] <n>
Specify limit; -100e6 to 100e6.
-2
√
[:DATA]?
Query lower limit.
√
:STATe <b>
Enable or disable Limit 2 test.
OFF
√
:STATe?
Query state of Limit 2 test.
√
:FAIL?
Return result of Limit 2 test; 0 (pass) or 1(fail).
√
:CLEar
Clear test results:
√
[:IMMediate]
Clear limit test results.
√
:AUTO <b>
Enable or disable clearing of limit test results
when a new trigger model cycle starts.
ON
√
:AUTO?
Query state of auto clear.
√
:IMMediate
Recalculate limit tests.
√
Table 14-2
CALibration command summary (user accessible)
Command
Description
Default
Parameter
Ref
SCPI
:CALibration
See Note.
:UNPRotected
Calibration accessible to operator:
:ACALibration
ACAL procedure:
:INITiate
Prepares 2182 for ACAL.
:STEP1
Performs a full ACAL (100V and 10mV).
:STEP2
Performs a limited ACAL (10mV only).
:DONE
Exit ACAL mode.
:TEMPerature?
Queries the internal temperature at the time of last
ACAL. If present internal temperature differs from
the last ACAL temperature by more than 1°C,
perform another ACAL.
Note: The above table only provides the commands to perform ACAL, which is a procedure to be performed by the
operator. The formal calibration is to be performed by qualified service personnel and is provided in the Model 2182
Service Manual.
Table 14-1
CALCulate command summary (cont.)
Summary of Contents for 2182
Page 1: ...www tek com keithley Model 2182 2182A Nanovoltmeter User s Manual 2182A 900 01 Rev B May 2017...
Page 18: ......
Page 22: ......
Page 23: ...1 Getting Started Getting Started...
Page 41: ...2 VoltageandTemperature Measurements Voltageand Temperature Measurements...
Page 68: ...2 28 Voltage and Temperature Measurements...
Page 69: ...3 Range Digits Rate andFilter Range Digits Rate andFilter...
Page 82: ...3 14 Range Digits Rate and Filter...
Page 83: ...4 Relative mX b and Percent Relative mX b andPercent...
Page 91: ...5 RatioandDelta Ratioand Delta...
Page 117: ...6 Buffer Buffer...
Page 123: ...7 Triggering Triggering...
Page 140: ...7 18 Triggering...
Page 141: ...8 Limits Limits...
Page 149: ...9 SteppingandScanning Steppingand Scanning...
Page 168: ...9 20 Stepping and Scanning...
Page 169: ...10 AnalogOutput Analog Output...
Page 175: ...11 RemoteOperation Remote Operation...
Page 205: ...12 CommonCommands Common Commands...
Page 221: ...13 SCPISignalOriented Measurement Commands SCPISignalOri entedMeasure ment Commands...
Page 225: ...14 SCPIReferenceTables SCPIRefer enceTables...
Page 239: ...15 AdditionalSCPI Commands Additional SCPICom mands...
Page 260: ...15 22 Additional SCPI Commands...
Page 261: ...A Specifications Specifications...
Page 263: ...B Statusand ErrorMessages StatusandError Messages...
Page 268: ...B 6 Status and Error Messages...
Page 269: ...C Measurement Considerations Measurement Consider ations...
Page 278: ...C 10 Measurement Considerations...
Page 279: ...D Model182Emulation Commands Model182 EmulationCom mands...
Page 284: ...D 6 Model 182 Emulation Commands...
Page 285: ...E Example Programs ExamplePro grams...
Page 293: ...F IEEE 488 BusOverview IEEE 488Bus Overview...
Page 307: ...G IEEE 488andSCPI ConformanceInformation IEEE 488and SCPIConform anceInforma tion...
Page 310: ...G 4 IEEE 488 and SCPI Conformance Information...
Page 311: ...H Measurement Queries Measurement Queries...
Page 316: ...H 6 Measurement Queries...
Page 317: ...I Delta PulseDelta and DifferentialConductance Delta Pulse Deltaand Dif ferentialCon ductance...