9-16
Stepping and Scanning
Set up 2400
Menu: Savesetup: Global: Reset: Bench
Meas: V
Source: I
Config Trig: ARM-LAYER: ARM-IN: IMMEDIATE
ARM-OUT : LINE: #3
: EVENTS: TRIG-LAYER-DONE= OFF
TRIG-LAYER: TRIGGER-IN: TRIGGER-LINK: #1: EVENT DETECT BYPASS
NEVER: TRIGGER IN EVENTS: SOURCE= ON all others off
TRIG-LAYER: TRIGGER-OUT: LINE: #2
: EVENTS :TRIGGER OUT EVENTS:
SOURCE = ON, all others off.
: COUNT 12
Config Sweep: TYPE: CUSTOM : #-POINTS: 12
: ADJUST-POINTS: see waveform
: COUNT: INFINITE
Speed: 0.01plc
Turn output: On
SWEEP :exit
TRIG: HALT
ON the 2182, enable SCAN (Note: Memory buffer annunciator comes on).
ON the 2400, enable SWEEP (Note: Arm annunciator comes on).
Press Trig on 2400.
After completion of the sweep, recall the data from the 2182 using the “TRACe” command.
To remove thermal EMFs from the readings, do the following math on the recalled data:
CH2 Rdg #1 = ( Buffer Rdg #1 - Buffer Rdg #5 ) / 2
CH2 Rdg #2 = ( Buffer Rdg #9 - Buffer Rdg #13 ) / 2
repeat.
CH1 Rdg_Pos #1 = ( Buffer Rdg #2 + Buffer Rdg #3 + Buffer Rdg #4 ) / 3
CH1 Rdg_Neg #1 = ( Buffer Rdg #6 + Buffer Rdg #7 + Buffer Rdg #8 ) / 3
CH1 Rdg#1 = ( CH1 Rdg_Pos #1 - CH1 Rdg_Neg #1 ) / 2
CH1 Rdg_Pos #2 = ( Buffer Rdg #10 + Buffer Rdg #11 + Buffer Rdg #12 ) / 3
CH1 Rdg_Neg #2 = ( Buffer Rdg #14 + Buffer Rdg #15 + Buffer Rdg #16 ) / 3
CH1 Rdg#2 = ( CH1 Rdg_Pos #2 - CH1 Rdg_Neg #2 ) / 2
repeat.
Summary of Contents for 2182
Page 1: ...www tek com keithley Model 2182 2182A Nanovoltmeter User s Manual 2182A 900 01 Rev B May 2017...
Page 18: ......
Page 22: ......
Page 23: ...1 Getting Started Getting Started...
Page 41: ...2 VoltageandTemperature Measurements Voltageand Temperature Measurements...
Page 68: ...2 28 Voltage and Temperature Measurements...
Page 69: ...3 Range Digits Rate andFilter Range Digits Rate andFilter...
Page 82: ...3 14 Range Digits Rate and Filter...
Page 83: ...4 Relative mX b and Percent Relative mX b andPercent...
Page 91: ...5 RatioandDelta Ratioand Delta...
Page 117: ...6 Buffer Buffer...
Page 123: ...7 Triggering Triggering...
Page 140: ...7 18 Triggering...
Page 141: ...8 Limits Limits...
Page 149: ...9 SteppingandScanning Steppingand Scanning...
Page 168: ...9 20 Stepping and Scanning...
Page 169: ...10 AnalogOutput Analog Output...
Page 175: ...11 RemoteOperation Remote Operation...
Page 205: ...12 CommonCommands Common Commands...
Page 221: ...13 SCPISignalOriented Measurement Commands SCPISignalOri entedMeasure ment Commands...
Page 225: ...14 SCPIReferenceTables SCPIRefer enceTables...
Page 239: ...15 AdditionalSCPI Commands Additional SCPICom mands...
Page 260: ...15 22 Additional SCPI Commands...
Page 261: ...A Specifications Specifications...
Page 263: ...B Statusand ErrorMessages StatusandError Messages...
Page 268: ...B 6 Status and Error Messages...
Page 269: ...C Measurement Considerations Measurement Consider ations...
Page 278: ...C 10 Measurement Considerations...
Page 279: ...D Model182Emulation Commands Model182 EmulationCom mands...
Page 284: ...D 6 Model 182 Emulation Commands...
Page 285: ...E Example Programs ExamplePro grams...
Page 293: ...F IEEE 488 BusOverview IEEE 488Bus Overview...
Page 307: ...G IEEE 488andSCPI ConformanceInformation IEEE 488and SCPIConform anceInforma tion...
Page 310: ...G 4 IEEE 488 and SCPI Conformance Information...
Page 311: ...H Measurement Queries Measurement Queries...
Page 316: ...H 6 Measurement Queries...
Page 317: ...I Delta PulseDelta and DifferentialConductance Delta Pulse Deltaand Dif ferentialCon ductance...