Model 182 Emulation Commands
D-5
NOTES:
1.
“
A” Commands
— The maximum number of characters for the A1 command string is
12. The A2 and A3 commands are not supported.
2.
“C” Commands
— Calibration commands are not supported by the 182 language. You
must use the SCPI language to calibrate the Model 2182 over the bus.
3.
“G” Commands
— The G4 through G7 commands are not supported by the 182
language. The Model 182 does not use time stamp.
4.
“H” Commands
— The H1 command is not supported by the 182 language.
5.
“I” Commands
— The minimum buffer size for the Model 2182 is 2, while the
minimum buffer size for the Model 182 is 1. Therefore, to set minimum buffer size, I1,0
is valid for the Model 182 and I1,2 is valid for the Model 2182. The I2 command
(circular buffer) is not supported.
6.
“J” Commands
— For the Model 2182, J1 uses the last reading as the Rel value
(whereas, the J1 command for the Model 182 uses the next reading as the Rel value). The
J3 command simply enables analog output relative and uses the present Rel value.
7.
“S” Commands
— The S3 command has been added to include a 1 second integration
period.
8.
“T” Commands
— The front panel TRIG key is always operational, even when T10X
is sent to disable all triggers.
9.
“U” Commands
:
•
U1
- For the Model 2182, the U1 error status word only supports the IDDC (Invalid
Device-Dependent Command) error (bit 0), the IDDCO (Invalid Device-Dependent
Command Option) error (bit 2), and the Uncalibrated Error (bit 8). All other bits
remain at zero.
•
U2
- The Model 2182 uses two processors, while the Model 182 uses three
processors. Therefore, the U2 command will be formatted to only provide revision
levels for the Front Panel Processor and the Main Processor.
10.
U11 and U14
— Not supported.
11.
“V” Commands
— The V1 command is not supported.
12.
“Z” Commands
— For the Model 2182, Z1 uses the last reading as the Rel value
(whereas, the Z1 command for the Model 182 uses the next reading as the Rel value).
The Z3 command simply enables reading relative and uses the present Rel value.
Summary of Contents for 2182
Page 1: ...www tek com keithley Model 2182 2182A Nanovoltmeter User s Manual 2182A 900 01 Rev B May 2017...
Page 18: ......
Page 22: ......
Page 23: ...1 Getting Started Getting Started...
Page 41: ...2 VoltageandTemperature Measurements Voltageand Temperature Measurements...
Page 68: ...2 28 Voltage and Temperature Measurements...
Page 69: ...3 Range Digits Rate andFilter Range Digits Rate andFilter...
Page 82: ...3 14 Range Digits Rate and Filter...
Page 83: ...4 Relative mX b and Percent Relative mX b andPercent...
Page 91: ...5 RatioandDelta Ratioand Delta...
Page 117: ...6 Buffer Buffer...
Page 123: ...7 Triggering Triggering...
Page 140: ...7 18 Triggering...
Page 141: ...8 Limits Limits...
Page 149: ...9 SteppingandScanning Steppingand Scanning...
Page 168: ...9 20 Stepping and Scanning...
Page 169: ...10 AnalogOutput Analog Output...
Page 175: ...11 RemoteOperation Remote Operation...
Page 205: ...12 CommonCommands Common Commands...
Page 221: ...13 SCPISignalOriented Measurement Commands SCPISignalOri entedMeasure ment Commands...
Page 225: ...14 SCPIReferenceTables SCPIRefer enceTables...
Page 239: ...15 AdditionalSCPI Commands Additional SCPICom mands...
Page 260: ...15 22 Additional SCPI Commands...
Page 261: ...A Specifications Specifications...
Page 263: ...B Statusand ErrorMessages StatusandError Messages...
Page 268: ...B 6 Status and Error Messages...
Page 269: ...C Measurement Considerations Measurement Consider ations...
Page 278: ...C 10 Measurement Considerations...
Page 279: ...D Model182Emulation Commands Model182 EmulationCom mands...
Page 284: ...D 6 Model 182 Emulation Commands...
Page 285: ...E Example Programs ExamplePro grams...
Page 293: ...F IEEE 488 BusOverview IEEE 488Bus Overview...
Page 307: ...G IEEE 488andSCPI ConformanceInformation IEEE 488and SCPIConform anceInforma tion...
Page 310: ...G 4 IEEE 488 and SCPI Conformance Information...
Page 311: ...H Measurement Queries Measurement Queries...
Page 316: ...H 6 Measurement Queries...
Page 317: ...I Delta PulseDelta and DifferentialConductance Delta Pulse Deltaand Dif ferentialCon ductance...