For this example, the Models 2182 and 7001/7002 are configured as follows:
Model 2182:
Factory defaults restored (accessed from SHIFT-SETUP)
External scanning, channels 1 - 8, no timer, 8 readings (accessed from SHIFT-CONFIG)
External triggers (accessed from EX TRIG)
Model 7001 or 7002:
Factory defaults restored
Scan list = 1!1-1!8,
Number of scans = 1
Channel spacing = TrigLink
To run the test and store readings in the Model 2182 with the unit set for external triggers,
press STEP or SCAN. The Model 2182 waits (with the asterisk annunciator lit) for an external
trigger from the Model 7001/7002.
Press STEP on the Model 7001/7002 to take it out of idle and start the scan. The scanner's
output pulse triggers the Model 2182 to take a reading, store it, and send a trigger pulse. The
following explanation on operation is referenced to the operation model shown in
Figure 7-8
Operation model for triggering example
Idle
Bypass
B
Wait for
Trigger Link
Trigger
Scan
Channel
C
Output
Trigger
Trigger
D
No
Scanned
8
Channels
?
Yes
7001or 7002
Make
Measurement
Made
8
Measurements
?
2182
Press STEP to start scan
A
Wait for
Trigger Link
Trigger
E
Output
Trigger
Trigger
F
No
Yes
Idle
7-10
Triggering
Summary of Contents for 2182
Page 1: ...www tek com keithley Model 2182 2182A Nanovoltmeter User s Manual 2182A 900 01 Rev B May 2017...
Page 18: ......
Page 22: ......
Page 23: ...1 Getting Started Getting Started...
Page 41: ...2 VoltageandTemperature Measurements Voltageand Temperature Measurements...
Page 68: ...2 28 Voltage and Temperature Measurements...
Page 69: ...3 Range Digits Rate andFilter Range Digits Rate andFilter...
Page 82: ...3 14 Range Digits Rate and Filter...
Page 83: ...4 Relative mX b and Percent Relative mX b andPercent...
Page 91: ...5 RatioandDelta Ratioand Delta...
Page 117: ...6 Buffer Buffer...
Page 123: ...7 Triggering Triggering...
Page 140: ...7 18 Triggering...
Page 141: ...8 Limits Limits...
Page 149: ...9 SteppingandScanning Steppingand Scanning...
Page 168: ...9 20 Stepping and Scanning...
Page 169: ...10 AnalogOutput Analog Output...
Page 175: ...11 RemoteOperation Remote Operation...
Page 205: ...12 CommonCommands Common Commands...
Page 221: ...13 SCPISignalOriented Measurement Commands SCPISignalOri entedMeasure ment Commands...
Page 225: ...14 SCPIReferenceTables SCPIRefer enceTables...
Page 239: ...15 AdditionalSCPI Commands Additional SCPICom mands...
Page 260: ...15 22 Additional SCPI Commands...
Page 261: ...A Specifications Specifications...
Page 263: ...B Statusand ErrorMessages StatusandError Messages...
Page 268: ...B 6 Status and Error Messages...
Page 269: ...C Measurement Considerations Measurement Consider ations...
Page 278: ...C 10 Measurement Considerations...
Page 279: ...D Model182Emulation Commands Model182 EmulationCom mands...
Page 284: ...D 6 Model 182 Emulation Commands...
Page 285: ...E Example Programs ExamplePro grams...
Page 293: ...F IEEE 488 BusOverview IEEE 488Bus Overview...
Page 307: ...G IEEE 488andSCPI ConformanceInformation IEEE 488and SCPIConform anceInforma tion...
Page 310: ...G 4 IEEE 488 and SCPI Conformance Information...
Page 311: ...H Measurement Queries Measurement Queries...
Page 316: ...H 6 Measurement Queries...
Page 317: ...I Delta PulseDelta and DifferentialConductance Delta Pulse Deltaand Dif ferentialCon ductance...