Remote Operation
11-11
Program Fragment
CALL TRANSMIT (“UNL LISTEN 7 LLO”, status%)‘ Lock out front panel.
CALL TRANSMIT (“UNL LISTEN 7 GTL”, status%)‘ Lock out front panel.
GTL (go to local)
Use the GTL command to put a remote mode instrument into local mode. The GTL command
also restores front panel key operation.
Program Fragment
CALL TRANSMIT (“MTA LISTEN 7 REN”, status%)‘ Place 2182 in remote.
CALL TRANSMIT (“UNL LISTEN 7 GTL”, status%)‘ Place 2182 in local mode
DCL (device clear)
Use the DCL command to clear the GPIB interface and return it to a known state. Note that
the DCL command is not an addressed command, so all instruments equipped to implement
DCL will do so simultaneously.
When the Model 2182 receives a DCL command, it clears the Input Buffer and Output Queue,
cancels deferred commands, and clears any command that prevents the processing of any other
device command. A DCL does not affect instrument settings and stored data.
Program fragment
CALL TRANSMIT (“DCL”, status%)‘ Clears all devices.
SDC (selective device clear)
The SDC command is an addressed command that performs essentially the same function as
the DCL command. However, since each device must be individually addressed, the SDC
command provides a method to clear only selected instruments instead of clearing all
instruments simultaneously, as is the case with DCL.
Program fragment
CALL TRANSMIT (“UNL LISTEN 7 SDC”, status%)‘ Clears 2182.
GET (group execute trigger)
GET is a GPIB trigger that is used as an event to control operation. The Model 2182 reacts to
this trigger if it is the programmed control source. The control source is programmed from the
SCPI TRIGger subsystem.
With the instrument programmed and waiting for a GPIB trigger, the following program
fragment will provide the GET:
Program fragment
CALL TRANSMIT (“UNL LISTEN 7 GET”, status%)
‘ Trigger 2182 from over
the bus.
When the command is executed, the trigger event for the 2182 occurs. Any other listeners
ignore the trigger.
Summary of Contents for 2182
Page 1: ...www tek com keithley Model 2182 2182A Nanovoltmeter User s Manual 2182A 900 01 Rev B May 2017...
Page 18: ......
Page 22: ......
Page 23: ...1 Getting Started Getting Started...
Page 41: ...2 VoltageandTemperature Measurements Voltageand Temperature Measurements...
Page 68: ...2 28 Voltage and Temperature Measurements...
Page 69: ...3 Range Digits Rate andFilter Range Digits Rate andFilter...
Page 82: ...3 14 Range Digits Rate and Filter...
Page 83: ...4 Relative mX b and Percent Relative mX b andPercent...
Page 91: ...5 RatioandDelta Ratioand Delta...
Page 117: ...6 Buffer Buffer...
Page 123: ...7 Triggering Triggering...
Page 140: ...7 18 Triggering...
Page 141: ...8 Limits Limits...
Page 149: ...9 SteppingandScanning Steppingand Scanning...
Page 168: ...9 20 Stepping and Scanning...
Page 169: ...10 AnalogOutput Analog Output...
Page 175: ...11 RemoteOperation Remote Operation...
Page 205: ...12 CommonCommands Common Commands...
Page 221: ...13 SCPISignalOriented Measurement Commands SCPISignalOri entedMeasure ment Commands...
Page 225: ...14 SCPIReferenceTables SCPIRefer enceTables...
Page 239: ...15 AdditionalSCPI Commands Additional SCPICom mands...
Page 260: ...15 22 Additional SCPI Commands...
Page 261: ...A Specifications Specifications...
Page 263: ...B Statusand ErrorMessages StatusandError Messages...
Page 268: ...B 6 Status and Error Messages...
Page 269: ...C Measurement Considerations Measurement Consider ations...
Page 278: ...C 10 Measurement Considerations...
Page 279: ...D Model182Emulation Commands Model182 EmulationCom mands...
Page 284: ...D 6 Model 182 Emulation Commands...
Page 285: ...E Example Programs ExamplePro grams...
Page 293: ...F IEEE 488 BusOverview IEEE 488Bus Overview...
Page 307: ...G IEEE 488andSCPI ConformanceInformation IEEE 488and SCPIConform anceInforma tion...
Page 310: ...G 4 IEEE 488 and SCPI Conformance Information...
Page 311: ...H Measurement Queries Measurement Queries...
Page 316: ...H 6 Measurement Queries...
Page 317: ...I Delta PulseDelta and DifferentialConductance Delta Pulse Deltaand Dif ferentialCon ductance...