15-10
Additional SCPI Commands
Operation Event Register:
Bit B0, Calibrating (Cal)
– Set bit indicates that the instrument is calibrating.
Bits B1 through B3
– Not used.
Bit B4, Measuring (Meas)
– Set bit indicates that the instrument is performing a
measurement.
Bit B5, Trigger Layer (Trig)
– Set bit indicates that the instrument is waiting in the Trigger
Layer of the Trigger Model.
Bits B6 and B7
– Not used.
Bit B8, Filter Settled (Filt)
– Set bit indicates that the filter has settled.
Bit B9
– Not used.
Bit B10, Idle
– Set bit indicates that the instrument is in the idle state.
Bits B11 through B15
– Not used.
Figure 15-6
Operation event register
Filt
B15 - B11
B10
B9
B8
B7,B6
B5
B4
B0
(2 )
5
32
0/1
Bit
Position
Event
Decimal
Weighting
Value
Value : 1 = Operation Event Set
0 = Operation Event Cleared
Events : Idle = Idle state of the 2182
Filt = Filter Settled
Trig = Trigger Layer
Meas = Measuring
Cal = Calibrating
Meas
(2 )
4
16
0/1
Idle
(2 )
10
1024
0/1
B3 - B1
Trig
(2 )
8
256
0/1
Cal
(2 )
0
1
0/1
Summary of Contents for 2182
Page 1: ...www tek com keithley Model 2182 2182A Nanovoltmeter User s Manual 2182A 900 01 Rev B May 2017...
Page 18: ......
Page 22: ......
Page 23: ...1 Getting Started Getting Started...
Page 41: ...2 VoltageandTemperature Measurements Voltageand Temperature Measurements...
Page 68: ...2 28 Voltage and Temperature Measurements...
Page 69: ...3 Range Digits Rate andFilter Range Digits Rate andFilter...
Page 82: ...3 14 Range Digits Rate and Filter...
Page 83: ...4 Relative mX b and Percent Relative mX b andPercent...
Page 91: ...5 RatioandDelta Ratioand Delta...
Page 117: ...6 Buffer Buffer...
Page 123: ...7 Triggering Triggering...
Page 140: ...7 18 Triggering...
Page 141: ...8 Limits Limits...
Page 149: ...9 SteppingandScanning Steppingand Scanning...
Page 168: ...9 20 Stepping and Scanning...
Page 169: ...10 AnalogOutput Analog Output...
Page 175: ...11 RemoteOperation Remote Operation...
Page 205: ...12 CommonCommands Common Commands...
Page 221: ...13 SCPISignalOriented Measurement Commands SCPISignalOri entedMeasure ment Commands...
Page 225: ...14 SCPIReferenceTables SCPIRefer enceTables...
Page 239: ...15 AdditionalSCPI Commands Additional SCPICom mands...
Page 260: ...15 22 Additional SCPI Commands...
Page 261: ...A Specifications Specifications...
Page 263: ...B Statusand ErrorMessages StatusandError Messages...
Page 268: ...B 6 Status and Error Messages...
Page 269: ...C Measurement Considerations Measurement Consider ations...
Page 278: ...C 10 Measurement Considerations...
Page 279: ...D Model182Emulation Commands Model182 EmulationCom mands...
Page 284: ...D 6 Model 182 Emulation Commands...
Page 285: ...E Example Programs ExamplePro grams...
Page 293: ...F IEEE 488 BusOverview IEEE 488Bus Overview...
Page 307: ...G IEEE 488andSCPI ConformanceInformation IEEE 488and SCPIConform anceInforma tion...
Page 310: ...G 4 IEEE 488 and SCPI Conformance Information...
Page 311: ...H Measurement Queries Measurement Queries...
Page 316: ...H 6 Measurement Queries...
Page 317: ...I Delta PulseDelta and DifferentialConductance Delta Pulse Deltaand Dif ferentialCon ductance...