5-12
Ratio and Delta
Step 4
Set up the SourceMeter to source current and measure voltage.
A.
On the SourceMeter, select Source I and Measure V.
B.
Select an appropriate current source range. For example, if your current reversal
values are going to ±1mA, select the 1mA source range.
C.
Press SPEED and select FAST. SourceMeter measurements are not used in this test,
but it must run as fast as possible to avoid synchronization problems with the
Model 2182.
Step 5
Configure the SourceMeter to perform a 2-point custom sweep.
A.
The menu to configure a sweep is accessed by pressing CONFIG and then SWEEP.
From the menu, select TYPE and then select CUSTOM.
Set the #-POINTS to two, and set ADJUST-POINTS (P0000 and P0001) to the
positive and negative current source values. For example, if the test requires 1mA,
set P0000 to +1mA and set P0001 to –1mA.
B.
Also from the sweep configuration menu, specify the number of 2-point sweeps to
perform:
Selecting INFINITE allows the SourceMeter to continuously source the current
reversal sweep. Select FINITE if you wish to perform a specific number of 2-point
sweeps.
Step 6
Return the Model 2182 to FACTory defaults.
Return the nanovoltmeter to its factory default conditions, by pressing RESTR and selecting
FACT.
Step 7
Configure the Model 2182 for Delta measurements.
A.
Press RATE to select 1 PLC (MED annunciator on) or 5 PLC (SLOW annunciator
on).
B.
Enable Delta measurements by pressing SHIFT and then V1-V2.
NOTE
For fast Delta measurements (2x speed), disable Front Autozero (SHIFT >
CONFIG > FRONT AUTOZERO N).
C.
Press EX TRIG to place the instrument in the external trigger mode. This will halt
measurements.
D.
If a longer settling time is required before performing each measurement, set a
manual delay from the Model 2182. Press SHIFT and then DELAY to select and set
delay.
NOTE
Do not set a delay on the SourceMeter, as this may adversely affect trigger
synchronization between the SourceMeter and the Model 2182.
Summary of Contents for 2182
Page 1: ...www tek com keithley Model 2182 2182A Nanovoltmeter User s Manual 2182A 900 01 Rev B May 2017...
Page 18: ......
Page 22: ......
Page 23: ...1 Getting Started Getting Started...
Page 41: ...2 VoltageandTemperature Measurements Voltageand Temperature Measurements...
Page 68: ...2 28 Voltage and Temperature Measurements...
Page 69: ...3 Range Digits Rate andFilter Range Digits Rate andFilter...
Page 82: ...3 14 Range Digits Rate and Filter...
Page 83: ...4 Relative mX b and Percent Relative mX b andPercent...
Page 91: ...5 RatioandDelta Ratioand Delta...
Page 117: ...6 Buffer Buffer...
Page 123: ...7 Triggering Triggering...
Page 140: ...7 18 Triggering...
Page 141: ...8 Limits Limits...
Page 149: ...9 SteppingandScanning Steppingand Scanning...
Page 168: ...9 20 Stepping and Scanning...
Page 169: ...10 AnalogOutput Analog Output...
Page 175: ...11 RemoteOperation Remote Operation...
Page 205: ...12 CommonCommands Common Commands...
Page 221: ...13 SCPISignalOriented Measurement Commands SCPISignalOri entedMeasure ment Commands...
Page 225: ...14 SCPIReferenceTables SCPIRefer enceTables...
Page 239: ...15 AdditionalSCPI Commands Additional SCPICom mands...
Page 260: ...15 22 Additional SCPI Commands...
Page 261: ...A Specifications Specifications...
Page 263: ...B Statusand ErrorMessages StatusandError Messages...
Page 268: ...B 6 Status and Error Messages...
Page 269: ...C Measurement Considerations Measurement Consider ations...
Page 278: ...C 10 Measurement Considerations...
Page 279: ...D Model182Emulation Commands Model182 EmulationCom mands...
Page 284: ...D 6 Model 182 Emulation Commands...
Page 285: ...E Example Programs ExamplePro grams...
Page 293: ...F IEEE 488 BusOverview IEEE 488Bus Overview...
Page 307: ...G IEEE 488andSCPI ConformanceInformation IEEE 488and SCPIConform anceInforma tion...
Page 310: ...G 4 IEEE 488 and SCPI Conformance Information...
Page 311: ...H Measurement Queries Measurement Queries...
Page 316: ...H 6 Measurement Queries...
Page 317: ...I Delta PulseDelta and DifferentialConductance Delta Pulse Deltaand Dif ferentialCon ductance...