5-22
Ratio and Delta
Superconductor Application #2 - fixed magnetic field
Another typical test on a superconductor sample (DUT) is to source an increasing-amplitude
current (I) through the DUT, while maintaining the magnetic field (H) at a fixed level. The I-V
curve in
shows that the measured voltage across the DUT remains at ~0V for low
currents (I). This is the flat portion of the curve where the DUT remains at 0
Ω
. At some point,
the voltage drop across the DUT will start increasing as current through the DUT increases. The
actual resistance of the DUT can be calculated at any current source point using Ohm’s law:
R = V/I
where: R is actual resistance of the DUT.
V is the measured voltage across the DUT.
I is the known current that flows through the DUT.
Figure 5-9
I-V Curve (Fixed H)
Such a test system is shown in
. A Keithley SourceMeter (Model 2400, 2410, or
2420) is used to source current through the DUT and two Model 2182s are used to provide
simultaneous voltage measurements.
Model 2182 #1 measures the voltage across a precision reference resistor (R
REF
) and stores
the readings in its buffer. These stored readings allow you to reference current amplitudes to the
voltage measurements of the DUT. Model 2182 #2 measures voltage across the DUT and stores
the readings in its buffer.
For example, assume you want to measure DUT voltage at current sweep values of 10µA,
20µA, and 50µA. When the sweep is started (10µA output), Model 2182 #1 measures 1mV
(10µA
×
100
Ω
= 1mV) and stores the reading in its buffer at location 1. At the same time,
Model 2182 #2 measures the DUT and stores that reading in its buffer at location 1. At the next
sweep point (20µA), Model 2182 #1 measures 2mV and stores the reading in its buffer at
location 2, and Model 2182 #2 measures the DUT and stores the reading in its buffer at location
2. At the last sweep point (50µA), Model 2182 #1 measures 5mV and stores the reading in its
buffer at location 3, and Model 2182 #2 measures the DUT and stores the reading in its buffer at
location 3.
Measure
V
Fixed H
Source I
Summary of Contents for 2182
Page 1: ...www tek com keithley Model 2182 2182A Nanovoltmeter User s Manual 2182A 900 01 Rev B May 2017...
Page 18: ......
Page 22: ......
Page 23: ...1 Getting Started Getting Started...
Page 41: ...2 VoltageandTemperature Measurements Voltageand Temperature Measurements...
Page 68: ...2 28 Voltage and Temperature Measurements...
Page 69: ...3 Range Digits Rate andFilter Range Digits Rate andFilter...
Page 82: ...3 14 Range Digits Rate and Filter...
Page 83: ...4 Relative mX b and Percent Relative mX b andPercent...
Page 91: ...5 RatioandDelta Ratioand Delta...
Page 117: ...6 Buffer Buffer...
Page 123: ...7 Triggering Triggering...
Page 140: ...7 18 Triggering...
Page 141: ...8 Limits Limits...
Page 149: ...9 SteppingandScanning Steppingand Scanning...
Page 168: ...9 20 Stepping and Scanning...
Page 169: ...10 AnalogOutput Analog Output...
Page 175: ...11 RemoteOperation Remote Operation...
Page 205: ...12 CommonCommands Common Commands...
Page 221: ...13 SCPISignalOriented Measurement Commands SCPISignalOri entedMeasure ment Commands...
Page 225: ...14 SCPIReferenceTables SCPIRefer enceTables...
Page 239: ...15 AdditionalSCPI Commands Additional SCPICom mands...
Page 260: ...15 22 Additional SCPI Commands...
Page 261: ...A Specifications Specifications...
Page 263: ...B Statusand ErrorMessages StatusandError Messages...
Page 268: ...B 6 Status and Error Messages...
Page 269: ...C Measurement Considerations Measurement Consider ations...
Page 278: ...C 10 Measurement Considerations...
Page 279: ...D Model182Emulation Commands Model182 EmulationCom mands...
Page 284: ...D 6 Model 182 Emulation Commands...
Page 285: ...E Example Programs ExamplePro grams...
Page 293: ...F IEEE 488 BusOverview IEEE 488Bus Overview...
Page 307: ...G IEEE 488andSCPI ConformanceInformation IEEE 488and SCPIConform anceInforma tion...
Page 310: ...G 4 IEEE 488 and SCPI Conformance Information...
Page 311: ...H Measurement Queries Measurement Queries...
Page 316: ...H 6 Measurement Queries...
Page 317: ...I Delta PulseDelta and DifferentialConductance Delta Pulse Deltaand Dif ferentialCon ductance...