11-4
Remote Operation
Interface selection and configuration procedures
When you select (enable) the GPIB interface, the RS-232 interface disables. Conversely,
selecting (enabling) the RS-232 interface disables the GPIB interface.
GPIB interface
The GPIB interface is selected and configured from the GPIB menu structure. From this
menu, you can enable or disable the GPIB interface, and check or change the following settings:
•
Primary address (0 to 30)
•
Language (SCPI or 182)
Perform the following steps to select and configure the GPIB interface:
NOTE
To retain a present GPIB setting, press ENTER with the setting displayed. You can exit
from the menu structure at any time by pressing EXIT.
1.
Press SHIFT and then GPIB to access the GPIB menu. The present state (on or off) of
the GPIB is displayed.
2.
To enable the GPIB interface:
A.
Place the cursor on the on/off selection by pressing the
key.
B.
Press the
or
key to toggle the selection to ON.
C.
Press ENTER. The present GPIB address is displayed.
3.
To change the GPIB address:
A.
Use the
,
,
, and
keys to display a valid address value.
B.
Press ENTER. The present language selection is displayed.
4.
To change the programming language:
A.
Place the cursor on the present language selection.
B.
Press the
or
key to toggle the selection.
C.
Press ENTER. The instrument returns to the normal display state.
RS-232 interface
The RS-232 interface is selected and configured from the RS-232 menu structure. From this
menu, you can enable or disable the RS-232 interface, and check or change the following
settings:
•
Baud rate (19.2k, 9600, 4800, 2400, 1200, 600, or 300)
•
Flow control (none or XonXoff)
•
Terminator (CR, LF, CRLF, or LFCR)
NOTE
” (located at the end of this section) for information
on the these settings and connections to the computer.
Summary of Contents for 2182
Page 1: ...www tek com keithley Model 2182 2182A Nanovoltmeter User s Manual 2182A 900 01 Rev B May 2017...
Page 18: ......
Page 22: ......
Page 23: ...1 Getting Started Getting Started...
Page 41: ...2 VoltageandTemperature Measurements Voltageand Temperature Measurements...
Page 68: ...2 28 Voltage and Temperature Measurements...
Page 69: ...3 Range Digits Rate andFilter Range Digits Rate andFilter...
Page 82: ...3 14 Range Digits Rate and Filter...
Page 83: ...4 Relative mX b and Percent Relative mX b andPercent...
Page 91: ...5 RatioandDelta Ratioand Delta...
Page 117: ...6 Buffer Buffer...
Page 123: ...7 Triggering Triggering...
Page 140: ...7 18 Triggering...
Page 141: ...8 Limits Limits...
Page 149: ...9 SteppingandScanning Steppingand Scanning...
Page 168: ...9 20 Stepping and Scanning...
Page 169: ...10 AnalogOutput Analog Output...
Page 175: ...11 RemoteOperation Remote Operation...
Page 205: ...12 CommonCommands Common Commands...
Page 221: ...13 SCPISignalOriented Measurement Commands SCPISignalOri entedMeasure ment Commands...
Page 225: ...14 SCPIReferenceTables SCPIRefer enceTables...
Page 239: ...15 AdditionalSCPI Commands Additional SCPICom mands...
Page 260: ...15 22 Additional SCPI Commands...
Page 261: ...A Specifications Specifications...
Page 263: ...B Statusand ErrorMessages StatusandError Messages...
Page 268: ...B 6 Status and Error Messages...
Page 269: ...C Measurement Considerations Measurement Consider ations...
Page 278: ...C 10 Measurement Considerations...
Page 279: ...D Model182Emulation Commands Model182 EmulationCom mands...
Page 284: ...D 6 Model 182 Emulation Commands...
Page 285: ...E Example Programs ExamplePro grams...
Page 293: ...F IEEE 488 BusOverview IEEE 488Bus Overview...
Page 307: ...G IEEE 488andSCPI ConformanceInformation IEEE 488and SCPIConform anceInforma tion...
Page 310: ...G 4 IEEE 488 and SCPI Conformance Information...
Page 311: ...H Measurement Queries Measurement Queries...
Page 316: ...H 6 Measurement Queries...
Page 317: ...I Delta PulseDelta and DifferentialConductance Delta Pulse Deltaand Dif ferentialCon ductance...