16 Calibrating Ci conductivity sensors
86
4
•
with temperature acquisition
A prerequisite is that temperature compensation was specified in
the calibration the presets.
Bring the temperature of the sample solution to the requested val-
ues of the reference and operation temperatures in succession.
The order does not matter. Acquisition of the individual values
takes place automatically.
•
without temperature acquisition
If temperature compensation was not specified in the calibration
presets, you must control acquisition of the value manually. Bring
the temperature of the sample solution first to the value of the ref-
erence temperature and confirm by pressing "OK".
Then proceed in the same manner for the operation temperature.
5
A Protocol summarizing the calibration values determined then ap-
pears. Acknowledge the protocol by pressing "OK".
Failed calibrations are canceled at this point and discarded.
6
Press "Yes" to accept the Calibration values determined and enter
the Calibration in the Calibration logbook.
Press "No" to discard the results.
Step
Action
requested
Temperature values
current temperature
value
Display after
acquisition of first
value
remaining
requested
Temperature value
Summary of Contents for AQUIS touch S
Page 8: ...Inhalt ...
Page 20: ...10 Configuration 20 Chapter 10 12 Calibration timer page 46 ...
Page 68: ...13 Calibrating redox sensors 68 ...
Page 78: ...15 Calibrating CR conductivity sensors 78 ...
Page 88: ...16 Calibrating Ci conductivity sensors 88 ...
Page 96: ...17 Calibrating universal inputs 96 ...
Page 106: ...18 Calibrating O DO sensors 106 ...
Page 146: ...21 Annex 146 ...
Page 147: ......
Page 148: ......
Page 149: ......