10 Configuration
46
10.12 Calibration timer
Every input for analysis measurands has its own calibration timer. Inputs for
temperature sensors do not have a calibration timer, since they do not need to
be calibrated. Calibration timers signal when sensor calibration is due via a cal-
ibration alarm. Once a particular input has been calibrated successfully, its cal-
ibration timer is reset. The signal for a due calibration can be generated, for
instance, by binary outputs with external external indicator lights or the alarm/
event list. The structure of the calibration timer configuration of sensors with
JUMO digiLine electronics differs from that of all other analysis sensors. In this
case, there is only the "Calibration interval" parameter. Observe the following
configuration table.
10.12.1
Configuration of the calibration timers
Open Calibration Timer Configuration Universal Inputs:
Device Menu > Configuration > Analog Inputs >
Universal Input 1 to 3 > Calibration Timer
Open Calibration Timer Configuration Analysis Inputs:
Device Menu > Configuration > Analog Inputs >
Analysis Input 1 to 4 > Calibration Timer
Open Calibration Timer Settings Digital Sensors:
Device Menu > Configuration > Digital Sensors > Digital Sensors 1 to 6 > Cali-
bration Timer
.
Configuration point Selection/
setting option
Explanation
Function
a
inactive
active
Activation/deactivation of the calibration timer
for an analog input
Alarm/Event list
off
event
alarm
Assignment of calibration timer timeout notifica-
tion to alarm list or event list
Alarm text
21 text characters
Text for the alarm/event list upon timeout of cali-
bration timers
Calibration interval
0 to 9999 days
Time from one calibration to the next. When a
calibration is due is indicated by the calibration
alarm on the JUMO AQUIS touch S.
a
For sensors with JUMO digiLine electronics, only the calibration interval can be set. With pH and redox
sensors with JUMO digiLine electronics, the calibration alarm is active automatically. The texts for alarm
and event list are preset.
Summary of Contents for AQUIS touch S
Page 8: ...Inhalt ...
Page 20: ...10 Configuration 20 Chapter 10 12 Calibration timer page 46 ...
Page 68: ...13 Calibrating redox sensors 68 ...
Page 78: ...15 Calibrating CR conductivity sensors 78 ...
Page 88: ...16 Calibrating Ci conductivity sensors 88 ...
Page 96: ...17 Calibrating universal inputs 96 ...
Page 106: ...18 Calibrating O DO sensors 106 ...
Page 146: ...21 Annex 146 ...
Page 147: ......
Page 148: ......
Page 149: ......