143
21 Annex
and therefore the supply voltage of the sensor may be lower.
For sensors of type JUMO digiLine pH/ORP/T, the calculation must be per-
formed with the peak values:
I
1
= I
S1
+ I
S2
+ I
S3
= 20 mA + 20 mA + 20 mA = 60 mA
I
2
= I
S2
+ I
S3
= 20 mA + 20 mA = 40 mA
I
3
= I
S3
= 20 mA
For sensors of type ecoLine O-DO/NTU, the highest peak value is used once
and the remaining sensors are considered with their average values. Example
for 1 x O-DO and 2 x NTU:
I
1
= I
S1
+ I
S2
+ I
S3
= 4 mA + 2 mA + 60 mA = 66 mA
I
2
= I
S2
+ I
S3
= 2 mA + 60 mA = 62 mA
I
3
= I
S3
= 60 mA
For further calculation it is assumed that the following sensors are used in the
bus structure shown above:
Sensor 1: JUMO digiLine pH (use peak value)
Sensor 2: ecoLine O-DO (use average value)
Sensor 3: ecoLine NTU (use peak value)
Thus, the following currents result:
I
1
= I
S1
+ I
S2
+ I
S3
= 20 mA + 4 mA + 60 mA = 84 mA =
0.084 A
I
2
= I
S2
+ I
S3
= 4 mA + 60 mA = 64 mA =
0.064 A
I
3
= I
S3
= 60 mA =
0.06 A
Step 2: Calculate the voltage drop on the individual line segments
The cable lengths of the line segments are 20 m each.
The voltage drop on a line segment is calculated in accordance with the follow-
ing formula:
U
x
=
ρ
× 2 × L
x
× I
x
/ A; where
ρ
= 1/56
Ω
mm
2
/m and A = 0.34 mm
2
In the above example this means:
U
1
=
ρ
× 2 × L
1
× I
1
/ A = 1/56
Ω
mm
2
/m × 2 ×
20 m × 0.084 A
/ 0.34 mm
2
=
0.177 V
Shown in simplified form:
U
1
= 1/56
Ω
× 2 ×
20 × 0.084 A
/ 0.34 =
0.177 V
U
2
= 1/56
Ω
× 2 ×
20 × 0.064 A
/ 0.34 =
0.135 V
U
3
= 1/56
Ω
× 2 ×
20 × 0.06 A
/ 0.34 =
0.126 V
Step 3: Calculate the voltage at the respective sensor
The value of the supply voltage at the respective sensor is given by the supply
voltage at the point of feed-in minus the sum of all the voltages which drop on
the line segments that are located between the point of feed-in and the sensor.
In the above example this means:
U
S1
= U
SV
- U
1
= 5.3 V - 0.177 V = 5.123 V
≈ 5.1 V
U
S2
= U
SV
- U
1
- U
2
= 5.3 V - 0.177 V - 0.135 V = 4.988 V
≈ 5.0 V
U
S3
= U
SV
- U
1
- U
2
- U
3
= 5.3 V - 0.177 V - 0.135 V - 0.126 V = 4.862 V
≈ 4.9 V
The required minimum voltage of the sensors is shown in the following table.
The voltage at sensor 1 (JUMO digiLine pH) is well above the minimum value
Sensor
Minimum voltage
JUMO digiLine pH/ORP/T
4.2 V
JUMO ecoLine O-DO
5 V
JUMO ecoLine NTU
5 V
Summary of Contents for AQUIS touch S
Page 8: ...Inhalt ...
Page 20: ...10 Configuration 20 Chapter 10 12 Calibration timer page 46 ...
Page 68: ...13 Calibrating redox sensors 68 ...
Page 78: ...15 Calibrating CR conductivity sensors 78 ...
Page 88: ...16 Calibrating Ci conductivity sensors 88 ...
Page 96: ...17 Calibrating universal inputs 96 ...
Page 106: ...18 Calibrating O DO sensors 106 ...
Page 146: ...21 Annex 146 ...
Page 147: ......
Page 148: ......
Page 149: ......