17 Calibrating universal inputs
92
•
free chlorine, pH/Temp.-compensated
In the universal inputs calibration default settings for the "free chlorine, pH/T-
compensated" operation mode, the slope calibration is enabled and preconfig-
ured as the only available calibration routine.
Additional calibration default settings for calibration of chlorine sensors are ex-
plained in the following table.
Sample screen:
Universal input calibra-
tion default settings in
the
"free chlorine, pH/T-
compensated" opera-
tion mode
Parameter
Possible settings
Explanation
Temperature com-
pensation
Selection from Analog selection
Temperature input for automatic acqui-
sition of the test/sample solution tem-
perature during the calibration
pH compensation
source
Selection from Analog selection
pH measurement input for automatic
sensing of the test/sample solution pH
value during the calibration
NOTE!
The "Calibration default settings" menu appears in the Device menu only if a
user with corresponding user rights is logged in.
⇨